Detection of simulated interstitial lung disease and catheters with selenium, storage phosphor, and film-based radiography

被引:34
作者
Bernhardt, TM
Otto, D
Reichel, G
Ludwig, K
Seifert, S
Kropf, S
Rapp-Bernhardt, U
机构
[1] Otto von Guericke Univ, Dept Diagnost Radiol, D-39120 Magdeburg, Germany
[2] Otto von Guericke Univ, Dept Biometr & Med Informat, D-39120 Magdeburg, Germany
关键词
lung; radiography; phantoms; comparative studies; digital; selenium detector; storage phosphor; receiver operating characteristic (ROC) curve;
D O I
10.1148/radiology.213.2.r99nv20445
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the diagnostic performance of storage phosphor and digital selenium radiography (DSR) with asymmetric and symmetric screen-film systems at different, speeds in the detection of simulated interstitial lung disease and catheters. MATERIALS AND METHODS: Patterns of simulated interstitial lung disease and catheters were superimposed over an anthropomorphic chest phantom. Hard-copy images were generated at DSR (200-, 400-, and 600-speed), storage phosphor radiography (200- and 400-speed), and asymmetric (400-speed) and symmetric (200- and 400-speed) screen-film imaging. Surface doses were measured, and receiver operating characteristic analyses were performed. RESULTS: No statistically significant differences were found between the detector systems with the same speeds for each interstitial pattern. Significantly poorer results were found at 600-speed DSR than at 200-speed DSR. Detection of catheters and nodules over high-attenuation areas was significantly worse with the symmetric screen-film system than with the other detectors. The surface dose with the DSR system, without a grid, was about 50% less than that of the other detector systems, with grids, tit the same speed. CONCLUSION: No significant difference was found in the diagnostic performance at DSR, storage phosphor radiography, and film-based radiography for simulated interstitial lung disease at corresponding speeds; there was a reduction in the surface dose of about 50% with the 400-speed DSR system.
引用
收藏
页码:445 / 454
页数:10
相关论文
共 22 条
[1]   TECHNICAL EVALUATION OF A DIGITAL CHEST RADIOGRAPHY SYSTEM THAT USES A SELENIUM DETECTOR [J].
CHOTAS, HG ;
FLOYD, CE ;
RAVIN, CE .
RADIOLOGY, 1995, 195 (01) :264-270
[2]  
DENCKER E, 1997, ROFO S, V166, P82
[3]   SELENIUM-BASED DIGITAL RADIOGRAPHY OF THE CHEST - RADIOLOGISTS PREFERENCE COMPARED WITH FILM-SCREEN RADIOGRAPHS [J].
FLOYD, CE ;
BAKER, JA ;
CHOTAS, HG ;
DELONG, DM ;
RAVIN, CE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (06) :1353-1358
[4]   Digital thoracic radiography using a selenium detector: Prospective comparison with conventional radiography [J].
Freund, M ;
Reuter, M ;
Palmie, S ;
Harder, E ;
Hutzelmann, A ;
Heller, M .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 166 (02) :101-107
[5]   DETECTION OF CENTRAL VENOUS CATHETERS WHEN USING STORAGE PHOSPHOR RADIOGRAPHY IN INTENSIVE-CARE RADIOLOGY [J].
GALANSKI, M ;
PROKOP, M ;
THORNS, E ;
OESTMANN, JW ;
REICHELT, S ;
HAUBITZ, B ;
MILBRADT, H ;
GRASER, A ;
VERNER, L ;
SCHAEFER, C .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1992, 156 (01) :68-72
[6]   VALUE OF THE ASYMMETRIC FILM-SCREEN SYSTEM INSIGHT HC IN CHEST IMAGING [J].
HAUSSLER, MD ;
LENZEN, H ;
RECKELS, C ;
PETERS, PE .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1994, 160 (03) :266-270
[7]   DIGITAL CHEST IMAGING WITH A SELENIUM DETECTOR - COMPARISON WITH CONVENTIONAL RADIOGRAPHY FOR VISUALIZATION OF SPECIFIC ANATOMIC REGIONS OF THE CHEST [J].
HEESEWIJK, HPM ;
NEITZEL, U ;
VANDERGRAAF, Y ;
DEVALOIS, JC ;
FELDBERG, MAM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (03) :535-540
[8]  
Kehler M, 1996, ACTA RADIOL, V37, P332
[9]  
KEHLER M, 1991, ACTA RADIOL, V32, P18
[10]   INTERPRETATION OF SUBTLE INTERSTITIAL LUNG ABNORMALITIES - CONVENTIONAL VERSUS FILM-DIGITIZED RADIOGRAPHY [J].
KIDO, S ;
IKEZOE, J ;
TAKEUCHI, N ;
KONDOH, H ;
JOHKOH, T ;
KOHNO, N ;
TOMIYAMA, N ;
YAMAGAMI, H ;
NAITO, H ;
ARISAWA, J ;
KOZUKA, T .
RADIOLOGY, 1994, 192 (01) :171-176