Adjustment method for mechanical Boston scientific corporation 30 MHz intravascular ultrasound catheters connected to a Clearview® console -: Mechanical 30 MHz IVUS Catheter Adjustment

被引:16
作者
Bruining, N
Hamers, R
Teo, TJ
de Feijter, PJ
Serruys, PW
Roelandt, JRTC
机构
[1] Erasmus MC, Rotterdam, Netherlands
[2] Boston Sci Corp, IVUS Technol Ctr, Fremont, CA USA
关键词
coronary disease; imaging; ultrasonics;
D O I
10.1023/B:CAIM.0000014046.63648.1c
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Intracoronary ultrasound (ICUS) is often used in studies evaluating new interventional techniques. It is important that quantitative measurements performed with various ICUS imaging equipment and materials are comparable. During evaluation of quantitative coronary ultrasound (QCU) software, it appeared that Boston Scientific Corporation (BSC) 30 MHz catheters connected to a Clearview(R) ultrasound console showed smaller dimensions of an in vitro phantom model than expected. In cooperation with the manufacturer the cause of this underestimation was determined, which is described in this paper, and the QCU software was extended with an adjustment. Evaluation was performed by performing in vitro measurements on a phantom model consisting of four highly accurate steel rings (perfect reflectors) with diameters of 2, 3, 4 and 5 mm. Relative differences (unadjusted) of the phantom were respectively: 15.92, 13.01, 10.10 and 12.23%. After applying the adjustment: -0.96, -1.84, -1.35 and -1.43%. In vivo measurements were performed on 24 randomly selected ICUS studies. These showed differences for not adjusted vs. adjusted measurements of lumen-, vessel- and plaque volumes of -10.1 +/- 1.5, -6.7 +/- 0.9 and -4.4 +/- 0.6%. An off-line adjustment formula was derived and applied on previous numerical QCU output data showing relative differences for lumen- and vessel volumes of 0.36 +/- 0.51 and 0.13 +/- 0.31%. 30 MHz BSC catheters connected to a Clearview(R) ultrasound console underestimate vessel dimensions. This can retrospectively be adjusted within QCU software as well as retrospectively on numerical QCU data using a mathematical model.
引用
收藏
页码:83 / 91
页数:9
相关论文
共 22 条
  • [11] A novel approach to quantitative analysis of intra vascular ultrasound images
    Hamers, R
    Bruining, N
    Knook, M
    Sabate, M
    Roelandt, JRTC
    [J]. COMPUTERS IN CARDIOLOGY 2001, VOL 28, 2001, 28 : 589 - 592
  • [12] Measurement of the speed of sound in ethanol/water mixtures
    Martin, K
    Spinks, D
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (02) : 289 - 291
  • [13] TARGET LESION CALCIFICATION IN CORONARY-ARTERY DISEASE - AN INTRAVASCULAR ULTRASOUND STUDY
    MINTZ, GS
    DOUEK, P
    PICHARD, AD
    KENT, KM
    SATLER, LF
    POPMA, JJ
    LEON, MB
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (05) : 1149 - 1155
  • [14] APPLICATION OF A NEW PHASED-ARRAY ULTRASOUND IMAGING CATHETER IN THE ASSESSMENT OF VASCULAR DIMENSIONS - INVIVO COMPARISON TO CINEANGIOGRAPHY
    NISSEN, SE
    GRINES, CL
    GURLEY, JC
    SUBLETT, K
    HAYNIE, D
    DIAZ, C
    BOOTH, DC
    DEMARIA, AN
    [J]. CIRCULATION, 1990, 81 (02) : 660 - 666
  • [15] Intravascular ultrasound - Novel pathophysiological insights and current clinical applications
    Nissen, SE
    Yock, P
    [J]. CIRCULATION, 2001, 103 (04) : 604 - 616
  • [16] Okabe T, 2000, J Invasive Cardiol, V12, P625
  • [17] 3-DIMENSIONAL RECONSTRUCTION OF INTRACORONARY ULTRASOUND IMAGES - RATIONALE, APPROACHES, PROBLEMS, AND DIRECTIONS
    ROELANDT, JRTC
    DIMARIO, C
    PANDIAN, NG
    WENGUANG, L
    KEANE, D
    SLAGER, CJ
    DEFEYTER, PJ
    SERRUYS, PW
    [J]. CIRCULATION, 1994, 90 (02) : 1044 - 1055
  • [18] Sustained suppression of neointimal proliferation by sirolimus-eluting stents - One-year angiographic and intravascular ultrasound follow-up
    Sousa, JE
    Costa, MA
    Abizaid, AC
    Rensing, BJ
    Abizaid, AS
    Tanajura, LF
    Kozuma, K
    Van Langenhove, G
    Sousa, AGMR
    Falotico, R
    Jaeger, J
    Popma, JJ
    Serruys, PW
    [J]. CIRCULATION, 2001, 104 (17) : 2007 - 2011
  • [19] Importance of calibration for diameter and area determination by intravascular ultrasound
    Stahr, P
    Rupprecht, HJ
    Voigtlander, T
    Kearney, P
    Erbel, R
    Koch, L
    Krass, S
    Brennecke, R
    Meyer, J
    [J]. INTERNATIONAL JOURNAL OF CARDIAC IMAGING, 1996, 12 (04): : 221 - 229
  • [20] The DICOM image formatting standard: its role in echocardiography and angiography
    Thomas, JD
    [J]. INTERNATIONAL JOURNAL OF CARDIAC IMAGING, 1998, 14 (Suppl 1): : 1 - 6