Photolithography was used to pattern hydrophobic design on common quantitative filter paper, based on which we successfully fabricated the paper-based micro-zone plates. In order to improve its protein-absorbing capability for further signal magnification, modified silicon dioxide (SiO2) micro-beads were deposited onto the surface of paper. Then indirect-ELISA for goat anti-rabbit IgG was applied on the 36-well plates. Taking enzyme linked goat anti-rabbit IgG as an example, loading SiO2 beads could remarkably increase the quantity of the absorbed protein by 20%-700%. When this method was used for goat anti-rabbit IgG detection, signals were magnified by 20%-150%, and the linear range of detection was from 3 x 10(-12) mmol/zone to 3 x 10(-8) mmol/zone under given conditions. In this manner, bio-reagent consumption was reduced to 3-5 mu L per zone, test time dropped to 25 min per plate and the cost of fabrication reduced to $1.1 per plate, which indicated a promising application foreground for new disease diagnosis devices.
引用
收藏
页码:20 / 24
页数:5
相关论文
共 22 条
[1]   Inkjet-printed paperfluidic immuno-chemical sensing device [J].
Abe, Koji ;
Kotera, Kaori ;
Suzuki, Koji ;
Citterio, Daniel .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 398 (02) :885-893
[2]   Low-cost printing of poly(dimethylsiloxane) barriers to define microchannels in paper [J].
Bruzewicz, Derek A. ;
Reches, Meital ;
Whitesides, George M. .
ANALYTICAL CHEMISTRY, 2008, 80 (09) :3387-3392
[3]   Paper Microzone Plates [J].
Carrilho, Emanuel ;
Phillips, Scott T. ;
Vella, Sarah J. ;
Martinez, Andres W. ;
Whitesides, George M. .
ANALYTICAL CHEMISTRY, 2009, 81 (15) :5990-5998
[4]   Electrochemical Detection in a Paper-Based Separation Device [J].
Carvalhal, Rafaela Fernanda ;
Kfouri, Marta Simao ;
de Oliveira Piazetta, Maria Helena ;
Gobbi, Angelo Luiz ;
Kubota, Lauro Tatsuo .
ANALYTICAL CHEMISTRY, 2010, 82 (03) :1162-1165
[5]   Paper-Based ELISA [J].
Cheng, Chao-Min ;
Martinez, Andres W. ;
Gong, Jinlong ;
Mace, Charles R. ;
Phillips, Scott T. ;
Carrilho, Emanuel ;
Mirica, Katherine A. ;
Whitesides, George M. .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2010, 49 (28) :4771-4774
[6]   A low-cost, simple, and rapid fabrication method for paper-based microfluidics using wax screen-printing [J].
Dungchai, Wijitar ;
Chailapakul, Orawon ;
Henry, Charles S. .
ANALYST, 2011, 136 (01) :77-82
[7]   Paper-based microfluidic devices for analysis of clinically relevant analytes present in urine and saliva [J].
Klasner, Scott A. ;
Price, Alexander K. ;
Hoeman, Kurt W. ;
Wilson, Rashaun S. ;
Bell, Kayla J. ;
Culbertson, Christopher T. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 397 (05) :1821-1829
[8]   Fabrication of paper-based microfluidic sensors by printing [J].
Li, Xu ;
Tian, Junfei ;
Garnier, Gil ;
Shen, Wei .
COLLOIDS AND SURFACES B-BIOINTERFACES, 2010, 76 (02) :564-570
[9]   Paper-Based Microfluidic Devices by Plasma Treatment [J].
Li, Xu ;
Tian, Junfei ;
Nguyen, Thanh ;
Shen, Wei .
ANALYTICAL CHEMISTRY, 2008, 80 (23) :9131-9134
[10]  
Liu XY, 2011, PROC IEEE MICR ELECT, P75, DOI 10.1109/MEMSYS.2011.5734365