Integration of multiple components in polystyrene-based microfluidic devices part II: cellular analysis

被引:21
作者
Anderson, Kari B. [2 ]
Halpin, Stephen T. [2 ]
Johnson, Alicia S. [1 ]
Martin, R. Scott [1 ]
Spence, Dana M. [2 ]
机构
[1] St Louis Univ, Dept Chem, St Louis, MO 63103 USA
[2] Michigan State Univ, Dept Chem, E Lansing, MI 48824 USA
关键词
CELLS; CULTURE; ELECTROPHORESIS; DOPAMINE; RELEASE; SYSTEMS;
D O I
10.1039/c2an36171j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In Part II of this series describing the use of polystyrene (PS) devices for microfluidic-based cellular assays: various cellular types and detection strategies are employed to determine three fundamental assays often associated with cells. Specifically, using either integrated electrochemical sensing or optical measurements with a standard multi-well plate reader, cellular uptake, production, or release of important cellular analytes are determined on a PS-based device. One experiment involved the fluorescence measurement of nitric oxide (NO) produced within an endothelial cell line following stimulation with ATP. The result was a four-fold increase in NO production (as compared to a control), with this receptor-based mechanism of NO production verifying the maintenance of cell receptors following immobilization onto the PS substrate. The ability to monitor cellular uptake was also demonstrated by optical determination of Ca2+ into endothelial cells following stimulation with the Ca2+ ionophore A20317. The result was a significant increase (42%) in the calcium uptake in the presence of the ionophore, as compared to a control (17%) (p < 0.05). Finally, the release of catecholamines from a dopaminergic cell line (PC 12 cells) was electrochemically monitored, with the electrodes being embedded into the PS-based device. The PC 12 cells had better adherence on the PS devices, as compared to use of PDMS. Potassium-stimulation resulted in the release of 114 +/- 11 mu M catecholamines, a significant increase (p < 0.05) over the release from cells that had been exposed to an inhibitor (reserpine, 20 +/- 2 mu M of catecholamines). The ability to successfully measure multiple analytes, generated in different means from various cells under investigation, suggests that PS may be a useful material for microfluidic device fabrication, especially considering the enhanced cell adhesion to PS, its enhanced rigidity/amenability to automation, and its ability to enable a wider range of analytes to be investigated, even analytes with a high degree of hydrophobicity.
引用
收藏
页码:137 / 143
页数:7
相关论文
共 19 条
[1]   Integration of on-chip peristaltic pumps and injection valves with microchip electrophoresis and electrochemical detection [J].
Bowen, Amanda L. ;
Martin, R. Scott .
ELECTROPHORESIS, 2010, 31 (15) :2534-2540
[2]   Shrink-Film Configurable Multiscale Wrinkles for Functional Alignment of Human Embryonic Stem Cells and their Cardiac Derivatives [J].
Chen, Aaron ;
Lieu, Deborah K. ;
Freschauf, Lauren ;
Lew, Valerie ;
Sharma, Himanshu ;
Wang, Jiaxian ;
Diep Nguyen ;
Karakikes, Ioannis ;
Hajjar, Roger J. ;
Gopinathan, Ajay ;
Botvinick, Elliot ;
Fowlkes, Charless C. ;
Li, Ronald A. ;
Khine, Michelle .
ADVANCED MATERIALS, 2011, 23 (48) :5785-+
[3]   Polystyrene-coated micropallets for culture and separation of primary muscle cells [J].
Detwiler, David A. ;
Dobes, Nicholas C. ;
Sims, Christopher E. ;
Kornegay, Joe N. ;
Allbritton, Nancy L. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 402 (03) :1083-1091
[4]   Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) [J].
Duffy, DC ;
McDonald, JC ;
Schueller, OJA ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4974-4984
[5]   Cells on chips [J].
El-Ali, Jamil ;
Sorger, Peter K. ;
Jensen, Klavs F. .
NATURE, 2006, 442 (7101) :403-411
[6]   Addressing a vascular endothelium array with blood components using underlying microfluidic channels [J].
Genes, Luiza I. ;
Tolan, Nicole V. ;
Hulvey, Matthew K. ;
Martin, R. Scott ;
Spence, Dana M. .
LAB ON A CHIP, 2007, 7 (10) :1256-1259
[7]   Direct Plate-Reader Measurement of Nitric Oxide Released from Hypoxic Erythrocytes Flowing through a Microfluidic Device [J].
Halpin, Stephen T. ;
Spence, Dana M. .
ANALYTICAL CHEMISTRY, 2010, 82 (17) :7492-7497
[8]  
Kovarik ML, 2012, ANAL CHEM, V84, P516, DOI [10.1021/ac202611x, 10.1021/ac3031543]
[9]   Voltammetric and pharmacological characterization of dopamine release from single exocytotic events at rat pheochromocytoma (PC12) cells [J].
Kozminski, KD ;
Gutman, DA ;
Davila, V ;
Sulzer, D ;
Ewing, AG .
ANALYTICAL CHEMISTRY, 1998, 70 (15) :3123-3130
[10]   Microchip-based integration of cell immobilization, electrophoresis, post-column derivatization, and fluorescence detection for monitoring the release of dopamine from PC 12 cells [J].
Li, Michelle W. ;
Martin, R. Scott .
ANALYST, 2008, 133 (10) :1358-1366