Radiation Dose Savings for Adult Pulmonary Embolus 64-MDCT Using Bismuth Breast Shields, Lower Peak Kilovoltage, and Automatic Tube Current Modulation

被引:85
作者
Hurwitz, Lynne M. [1 ]
Yoshizumi, Terry T. [1 ]
Goodman, Philip C. [1 ]
Nelson, Rendon C. [1 ]
Toncheva, Greta [1 ]
Nguyen, Giao B. [1 ]
Lowry, Carolyn [1 ]
Anderson-Evans, Colin [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
关键词
bismuth breast shields; chest imaging protocol; MDCT angiography; pulmonary embolus; radiation dose; IMAGE QUALITY; COMPUTED-TOMOGRAPHY; CHEST CT; ANGIOGRAPHY; EXPOSURE; PROTOCOLS; PHANTOM; MDCT;
D O I
10.2214/AJR.08.1066
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
OBJECTIVE. The purpose of this study was to assess whether radiation dose savings using a lower peak kilovoltage (kVp) setting, bismuth breast shields, and automatic tube current modulation could be achieved while preserving the image quality of MDCT scans obtained to assess for pulmonary embolus (PE). MATERIALS AND METHODS. CT angiography (CTA) examinations were performed to assess for the presence or absence of pulmonary artery emboli using a 64-MDCT scanner with automatic tube current modulation (noise level = 10 HU), two kVp settings (120 and 140 kVp), and bismuth breast shields. Absorbed organ doses were measured using anthropomorphic phantoms and metal oxide semiconductor field effect transistor (MOSFET) detectors. Image quality was assessed quantitatively as well as qualitatively in various anatomic sites of the thorax. RESULTS. Using a lower kVp (120 vs 140 kVp) and automatic tube current modulation resulted in a dose savings of 27% to the breast and 47% to the lungs. The use of a lower kVp (120 kVp), automatic tube current modulation, and bismuth shields placed directly on the anterior chest wall reduced absorbed breast and lung doses by 55% and 45%, respectively. Qualitative assessment of the images showed no change in image quality of the lungs and mediastinum when using a lower kVp, bismuth shields, or both. CONCLUSION. The use of bismuth breast shields together with a lower kVp and automatic tube current modulation will reduce the absorbed radiation dose to the breast and lungs without degradation of image quality to the organs of the thorax for CTA detection of PE.
引用
收藏
页码:244 / 253
页数:10
相关论文
共 20 条
[1]
MDCT of the coronary arteries: Feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose [J].
Abada, Hicham T. ;
Larchez, Christophe ;
Daoud, Beatrice ;
Sigal-Cinqualbre, Anne ;
Paul, Jean-Francois .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :S387-S390
[2]
COURSEY C, 2008, AJR, V190, P244
[3]
Pediatric chest MDCT using tube current modulation: Effect on radiation dose with breast shielding [J].
Coursey, Courtney ;
Frush, Donald P. ;
Yoshizumi, Terry ;
Toncheva, Greta ;
Nguyen, Giao ;
Greenberg, S. Bruce .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (01) :W54-W61
[4]
Low-dose ECG-gated 64-slices helical CT angiography of the chest: evaluation of image quality in 105 patients [J].
d'Agostino, A. G. ;
Remy-Jardin, M. ;
Khalil, C. ;
Delannoy-Deken, V. ;
Flohr, T. ;
Duhamel, A. ;
Remy, J. .
EUROPEAN RADIOLOGY, 2006, 16 (10) :2137-2146
[5]
Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality [J].
Geleijns, J. ;
Salvado Artells, M. ;
Veldkamp, W. J. H. ;
Lopez Tortosa, M. ;
Calzado Cantera, A. .
EUROPEAN RADIOLOGY, 2006, 16 (10) :2334-2340
[6]
Image quality and radiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol: prospective randomized study [J].
Heyer, Christoph M. ;
Mohr, Patrick S. ;
Lemburg, Stefan P. ;
Peters, Soeren A. ;
Nicolas, Volkmar .
RADIOLOGY, 2007, 245 (02) :577-583
[7]
Radiation dose from contemporary cardiothoracic multidetector CT Protocols with an anthropomorphic female phantom: Implications for cancer induction [J].
Hurwitz, Lynne M. ;
Reiman, Robert E. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Toncheva, Greta ;
Nguyen, Giao ;
Lowry, Carolyn .
RADIOLOGY, 2007, 245 (03) :742-750
[8]
Effective dose determination using an anthropomorphic phantom and metal oxide semiconductor field effect transistor technology for clinical adult body multidetector array computed tomography protocols [J].
Hurwitz, Lynne M. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Frush, Donald P. ;
Nguyen, Giao ;
Toncheva, Greta ;
Lowry, Carolyn .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2007, 31 (04) :544-549
[9]
Techniques and applications of automatic tube current modulation for CT [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Schmidt, B ;
Westerman, BL ;
Morgan, HT ;
Saini, S .
RADIOLOGY, 2004, 233 (03) :649-657
[10]
Reduced-dose CT: Effect on reader evaluation in detection of pulmonary embolism [J].
MacKenzie, John D. ;
Nazario-Larrieu, Javier ;
Cai, Tianxi ;
Ledbetter, M. Stephen ;
Duran-Mendicuti, Maria Alejandra ;
Judy, Philip F. ;
Rybicki, Frank J. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (06) :1371-1379