5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets -: Two randomized studies using radiostereometric analysis

被引:144
作者
Digas, Georgios [1 ]
Karrholm, Johan [1 ]
Thanner, Jonas [1 ]
Herberts, Peter [1 ]
机构
[1] Univ Gothenburg, Sahlgrens Univ Hosp, Cent Hosp Molndal, Inst Surg Sci,Dept Orthoped, Gothenburg, Sweden
关键词
D O I
10.1080/17453670710014518
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Laboratory tests and early clinical studies have shown that highly cross-linked polyethylene (PE) markedly improves wear resistance compared to conventional PE. We evaluated this type of PE in two randomized clinical studies using radiostereometric analysis (RSA). The 2- and 3-year follow-up of these studies have already been reported. We found a lower penetration rate for the highly cross-linked PE than for conventional PE. We now report the outcome after 5 years. Patients and methods 60 patients (61 hips) with a median age of 55 years were randomized to receive either highly cross-linked PE (Durasul) or conventional cemented all-polyethylene sockets of the same design. 55 patients (56 hips) were followed for 5 years. In the second study, 32 patients (64 hips) with a median age of 48 years and with bilateral primary or secondary osteoarthritis of the hip had hybrid total hip arthroplasty with liners made of highly cross-linked PE (Longevity) on one side and conventional PE on the other. 23 patients in this study have passed the 5-year follow-up. Results Bedding-in and creep for the Durasul all-PE was reached by 2 years, and by I year for the Longevity liners. In both control groups with conventional PE, this was reached by 6 months. The steady-state wear rate was close to zero in the two study groups and 0.06 mm/year for the conventional PE in the two control groups. Interpretation We found that use of highly crosslinked PE instead of conventional PE reduced the wear by more than 95%, which supports continued use of this type of PE in young, active patients.
引用
收藏
页码:746 / 754
页数:9
相关论文
共 24 条
[1]
Standing versus supine radiographs in RSA evaluation of femoral head penetration [J].
Bragdon, Charles R. ;
Thanner, Jonas ;
Greene, Meridith E. ;
Malchau, Henrik ;
Digas, Georgios ;
Harris, William H. ;
Karrholm, Johan .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (448) :46-51
[2]
Five-year experience with Crossfire® highly cross-linked polyethylene [J].
D'Antonio, JA ;
Manley, MT ;
Capello, WN ;
Bierbaum, BE ;
Ramakrishnan, R ;
Naughton, M ;
Sutton, K .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (441) :143-150
[3]
Highly cross-linked polyethylene in cemented THA -: Randomized study of 61 hips [J].
Digas, G ;
Kärrholm, J ;
Thanner, J ;
Malchau, H ;
Herberts, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :126-138
[4]
The Otto Aufranc Award -: Highly cross-linked polyethylene in Total Hip Arthroplasty -: Randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis [J].
Digas, G ;
Kärrholm, J ;
Thanner, J ;
Malchau, H ;
Herberts, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :6-16
[5]
Clinical performance of a durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years [J].
Dorr, LD ;
Wan, ZN ;
Shahrdar, C ;
Sirianni, L ;
Boutary, M ;
Yun, A .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08) :1816-1821
[6]
A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[7]
A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners [J].
Engh, C. Anderson, Jr. ;
Stepniewski, Adam S. ;
Ginn, Stuart D. ;
Beykirch, Sarah E. ;
Sychterz-Terefenko, Christi J. ;
Hopper, Robert H., Jr. ;
Engh, Charles A. .
JOURNAL OF ARTHROPLASTY, 2006, 21 (06) :17-25
[8]
Effect of polyethylene wear on osteolysis in cementless primary total hip arthroplasty - Minimal 5-year follow-up study [J].
Han, CD ;
Choe, WS ;
Yoo, JH .
JOURNAL OF ARTHROPLASTY, 1999, 14 (06) :714-723
[9]
Harris WH, 2001, CLIN ORTHOP RELAT R, P66
[10]
Short-term in vivo wear of cross-linked polyethylene [J].
Heisel, C ;
Silva, M ;
dela Rosa, MA ;
Schmalzried, TP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (04) :748-751