Determination of the presampled MTF in computed tomography

被引:148
作者
Boone, JM [1 ]
机构
[1] Univ Calif Davis, Med Ctr, Dept Radiol, Res Imaging Ctr, Sacramento, CA 95817 USA
关键词
computed tomography; presampled MTF; line spread function; quality assurance; x-ray CT;
D O I
10.1118/1.1350438
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A technique for measuring the presampled MTF in CT scanners is described. The technique uses a simple phantom consisting of similar to0.050 mm aluminum foil sandwiched by flat plastic or tissue equivalent slabs. The aluminum foil is slightly angled with respect to the reconstruction matrix, and CT images are acquired. The acquired CT image yields an angled slit image that can be used to synthesize the presampled line spread function (LSF). The presampled MTF is calculated from the presampled LSF. The technique is a direct extension of that proposed by Fujita et al. [IEEE Trans. Med. Imaging 11, 34-39 (1993)] for MTF calculation on digital radiography images. While the MTF in clinical CT scanners often reaches negligible amplitude below the Nyquist frequency, the technique is easy to implement, requires inexpensive materials, is robust to aliasing, and is more resilient to noise due to greater data averaging than conventional PSF-integration techniques. Use of the proposed technique is illustrated on a clinical multiple detector array scanner, and MTFs are shown for several common reconstruction kernels. It is likely that the proposed technique would be useful for all tomographic imaging systems, including single photon emission computed tomography, positron emission tomography, magnetic resonance imaging and ultrasound scanners. (C) 2001 American Association of Physicists in Medicine. [DOI: 10.1118/1.1350438].
引用
收藏
页码:356 / 360
页数:5
相关论文
共 22 条
  • [1] Quantification of periapical bone destruction in mice by micro-computed tomography
    Balto, K
    Müller, R
    Carrington, DC
    Dobeck, J
    Stashenko, P
    [J]. JOURNAL OF DENTAL RESEARCH, 2000, 79 (01) : 35 - 40
  • [2] BISCHOF C J, 1977, Medical Physics (Woodbury), V4, P163, DOI 10.1118/1.594305
  • [3] Sinusoidal modulation analysis for optical system MTF measurements
    Boone, JM
    Yu, T
    Seibert, JA
    [J]. MEDICAL PHYSICS, 1996, 23 (12) : 1955 - 1963
  • [4] A multiple detector array helical x-ray microtomography system for specimen imaging
    Corrigan, NM
    Chavez, AE
    Wisner, ER
    Boone, JM
    [J]. MEDICAL PHYSICS, 1999, 26 (08) : 1708 - 1713
  • [5] Dainty JC., 1974, IMAGE SCI
  • [6] EFFECTS OF UNDERSAMPLING ON THE PROPER INTERPRETATION OF MODULATION TRANSFER-FUNCTION, NOISE POWER SPECTRA, AND NOISE EQUIVALENT QUANTA OF DIGITAL IMAGING-SYSTEMS
    DOBBINS, JT
    [J]. MEDICAL PHYSICS, 1995, 22 (02) : 171 - 181
  • [7] A PRACTICAL METHOD TO MEASURE THE MTF OF CT SCANNERS
    DROEGE, RT
    MORIN, RL
    [J]. MEDICAL PHYSICS, 1982, 9 (05) : 758 - 759
  • [8] HIGH-CONTRAST RESOLUTION OF CT IMAGES FOR BONE-STRUCTURE ANALYSIS
    DURAND, EP
    RUEGSEGGER, P
    [J]. MEDICAL PHYSICS, 1992, 19 (03) : 569 - 573
  • [9] A SIMPLE METHOD FOR DETERMINING THE MODULATION TRANSFER-FUNCTION IN DIGITAL RADIOGRAPHY
    FUJITA, H
    TSAI, DY
    ITOH, T
    DOI, K
    MORISHITA, J
    UEDA, K
    OHTSUKA, A
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) : 34 - 39
  • [10] HUDA W, 1987, J CAN ASSOC RADIOL, V38, P122