Final incision size after implantation of a hydrophobic acrylic aspheric intraocular lens: New motorized injector versus standard manual injector

被引:25
作者
Allen, David [1 ]
Habib, Maged [1 ]
Steel, David [1 ]
机构
[1] Sunderland Eye Infirm, Sunderland SR2 9HP, Tyne & Wear, England
关键词
MICROINCISION CATARACT-SURGERY; POSTOPERATIVE ENDOPHTHALMITIS; COHERENCE TOMOGRAPHY; INDUCED ASTIGMATISM; PHACOEMULSIFICATION; MONARCH;
D O I
10.1016/j.jcrs.2011.08.038
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
PURPOSE: To assess the impact on incision size of IOL implantation using a new motorized intraocular lens (IOL) injector versus a standard manual injector. SETTING: Cataract Treatment Centre, Sunderland Eye Infirmary, Sunderland, United Kingdom. DESIGN: Comparative case series. METHODS: All patients received an Acrysof SN60WF IOL implanted using a D cartridge and an Autosert motorized injector at fast speed, an Autosert motorized injector at slow speed, or a manual Monarch injector. Each group had a range of preimplantation incision sizes (1.9 mm, 2.0 mm, 2.1 mm, 2.2 mm, 2.3 mm). Incision gauges were used to measure the incision width before and immediately after IOL implantation. RESULTS: The study recruited 256 patients. All incisions that were 1.8 mm at the commencement of surgery increased in size by the end of irrigation/aspiration. The motorized injector used with a fast speed (4.4 mm/sec) caused significantly less incision enlargement than the manual injector for all preimplantation incision sizes tested (P<.02). For 4 of the 5 preimplantation incision size subgroups, the motorized injector used at slow speed (1.5 mm/sec) produced less incision stretch than the manual injector, although the difference did not reach statistical significance. CONCLUSION: The motorized injector was easy to use, and its use with an injection speed of 4.4 mm/sec caused significantly less incision enlargement than the manual injector during IOL implantation. Financial Disclosure: Drs. Allen and Steele have received travel and lodging reimbursement from Alcon Laboratories, Inc., Fort Worth, Texas, USA, which also provided the motorized implantation device for evaluation purposes. Dr. Habib has no financial or proprietary interest in any material or method mentioned.
引用
收藏
页码:249 / 255
页数:7
相关论文
共 20 条
[2]
Clear corneal incision architecture in the immediate postoperative period evaluated using optical coherence tomography [J].
Calladine, Daniel ;
Packard, Richard .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (08) :1429-1435
[3]
Aspheric microincision intraocular lens implantation with biaxial microincision cataract surgery: Efficacy and reliability [J].
Can, Izzet ;
Takmaz, Tamer ;
Bayhan, Hasan Ali ;
Ceran, Basak Bostanci .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (11) :1905-1911
[4]
Corneal incision quality: Microincision cataract surgery versus microcoaxial phacoemulsification [J].
Elkady, Bassam ;
Pinero, David ;
Alio, Jorge L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (03) :466-474
[5]
Profile of clear corneal cataract incisions demonstrated by ocular coherence tomography [J].
Fine, I. Howard ;
Hoffman, Richard S. ;
Packer, Mark .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (01) :94-97
[6]
Postoperative corneal shape changes: Microincision versus small-incision coaxial cataract surgery [J].
Hayashi, Ken ;
Yoshida, Motoaki ;
Hayashi, Hideyuki .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (02) :233-239
[7]
COMPARISON OF THE INDUCED ASTIGMATISM AFTER TEMPORAL CLEAR CORNEAL TUNNEL INCISIONS OF DIFFERENT SIZES [J].
KOHNEN, T ;
DICK, B ;
JACOBI, KW .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1995, 21 (04) :417-424
[8]
Incision sizes before and after implantation of 6-mm optic foldable intraocular lenses using monarch and unfolder injector systems [J].
Kohnen, T ;
Kasper, T .
OPHTHALMOLOGY, 2005, 112 (01) :58-66
[9]
Incision sizes for foldable intraocular lenses [J].
Kohnen, T ;
Lambert, RJ ;
Koch, DD .
OPHTHALMOLOGY, 1997, 104 (08) :1277-1286
[10]
Incision sizes before and after implantation of SN60WF intraocular lenses using the Monarch injector system with C and D cartridges [J].
Kohnen, Thomas ;
Klaproth, Oliver K. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (10) :1748-1753