The variation in output of symmetric, asymmetric and irregularly shaped wedged radiotherapy fields

被引:14
作者
Hounsell, AR
Wilkinson, JM
机构
[1] North Western Medical Physics, Christie Hospital, Holt Radium Institute
关键词
D O I
10.1088/0031-9155/41/10/021
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A model for calculating the variation in output of symmetric, asymmetric and irregularly shaped wedged radiotherapy fields is presented. The variation in output from the treatment head when a wedge is used is calculated by dividing the output into a primary component and one due to scattered radiation. The scatter component is then further sub-divided into contributions from elements which have a 1 cm x 1 cm cross-sectional area at the isocentre. The scatter from each element is determined as the contribution from the head scatter component modified by the presence of the wedge and a contribution due to additional scattered radiation from the wedge. The relative intensity of the scattered radiation from the wedge is modelled using a simple first scatter approximation. In this approximation the magnitude of the scatter is given by a t exp(-mu t) function where t is the thickness of the wedge for the selected element. The magnitude of the primary component and the relative intensity of scatter from each element are then obtained by an iterative fit to measured data. The technique has been applied to two different internally mounted wedge designs, for a standard treatment head, two asymmetric treatment heads and two similar multileaf collimators, over a range of energies between 4 and 20 MV. Calculations agree with measured values over a range of field sizes and shapes to within 1.5%.
引用
收藏
页码:2155 / 2172
页数:18
相关论文
共 30 条
[1]   MODELING TRANSMISSION AND SCATTER FOR PHOTON-BEAM ATTENUATORS [J].
AHNESJO, A ;
WEBER, L ;
NILSSON, P .
MEDICAL PHYSICS, 1995, 22 (11) :1711-1720
[2]   COLLIMATOR SCATTER IN PHOTON THERAPY BEAMS [J].
AHNESJO, A .
MEDICAL PHYSICS, 1995, 22 (03) :267-278
[3]   ANALYTIC MODELING OF PHOTON SCATTER FROM FLATTENING FILTERS IN PHOTON THERAPY BEAMS [J].
AHNESJO, A .
MEDICAL PHYSICS, 1994, 21 (08) :1227-1235
[4]   WEDGE TRANSMISSION DATA - EFFECT OF BACKSCATTER INTO MONITOR CHAMBER [J].
ATHERTON, L .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (02) :451-452
[5]   THE RELATION BETWEEN WEDGE FACTORS IN AIR AND WATER [J].
BARDEROMA, RD ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1994, 21 (07) :1043-1047
[6]   MONTE-CARLO STUDY OF ACCELERATOR HEAD SCATTER [J].
CHANEY, EL ;
CULLIP, TJ ;
GABRIEL, TA .
MEDICAL PHYSICS, 1994, 21 (09) :1383-1390
[7]   GAMMA-RAY ABSORPTION COEFFICIENTS [J].
DAVISSON, CM ;
EVANS, RD .
REVIEWS OF MODERN PHYSICS, 1952, 24 (02) :79-107
[8]   ON THE FIELD-SIZE DEPENDENCE OF RELATIVE OUTPUT FROM A LINEAR-ACCELERATOR [J].
DUNSCOMBE, PB ;
NIEMINEN, JM .
MEDICAL PHYSICS, 1992, 19 (06) :1441-1444
[9]   WEDGE FACTOR CONSTITUENTS OF HIGH-ENERGY PHOTON BEAMS - FIELD SIZE AND DEPTH DEPENDENCE [J].
HEUKELOM, S ;
LANSON, JH ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1994, 30 (01) :66-73
[10]   WEDGE FACTOR CONSTITUENTS OF HIGH-ENERGY PHOTON BEAMS - HEAD AND PHANTOM SCATTER DOSE COMPONENTS [J].
HEUKELOM, S ;
LANSON, JH ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1994, 32 (01) :73-83