Alumina-on-alumina hip implants A WEAR STUDY OF RETRIEVED COMPONENTS

被引:22
作者
Affatato, S. [1 ]
Traina, F. [1 ]
De Fine, M. [1 ]
Carmignato, S. [1 ]
Toni, A. [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2012年 / 94B卷 / 01期
关键词
CERAMIC BALL HEADS; IN-VITRO; ZRO2; FRACTION; SIMULATOR;
D O I
10.1302/0301-620X.94B1.26956
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Alumina-alumina bearings are among the most resistant to wear in total hip replacement. Examination of their surfaces is one way of comparing damage caused by wear of hip joints simulated in vitro to that seen in explanted bearings. The aim of this study was to determine whether second-generation ceramic bearings exhibited a better pattern of wear than those reported in the literature for first-generation bearings. We considered both macro-and microscopic findings. We found that long-term alumina wear in association with a loose acetabular component could be categorised into three groups. Of 20 specimens, four had 'low wear', eight 'crescent wear' and eight 'severe wear', which was characterised by a change in the physical shape of the bearing and a loss of volume. This suggests that the wear in alumina-alumina bearings in association with a loose acetabular component may be variable in pattern, and may explain, in part, why the wear of a ceramic head in vivo may be greater than that seen after in vitro testing.
引用
收藏
页码:37 / 42
页数:6
相关论文
共 17 条
[1]
The predictive power of surface profile parameters on the amount of wear measured in vitro on metal-on-polyethylene artificial hip joints [J].
Affatato, S ;
Bersaglia, G ;
Yin, JQ ;
Traina, F ;
Toni, A ;
Viceconti, M .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H3) :457-464
[2]
Is Ceramic-On-Ceramic Squeaking Phenomenon Reproducible In Vitro? A Long-Term Simulator Study Under Severe Conditions [J].
Affatato, S. ;
Traina, F. ;
Mazzega-Fabbro, C. ;
Sergo, V. ;
Viceconti, M. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 91B (01) :264-271
[3]
Mixed-oxides prosthetic ceramic ball heads.: Part 2:: Effect of the ZrO2 fraction on the wear of ceramic on ceramic joints [J].
Affatato, S ;
Testoni, M ;
Cacciari, GL ;
Toni, A .
BIOMATERIALS, 1999, 20 (20) :1925-1929
[4]
Mixed oxides prosthetic ceramic ball heads.: Part 3:: effect of the ZrO2 fraction on the wear of ceramic on ceramic hip joint prostheses.: A long-term in vitro wear study [J].
Affatato, S ;
Goldoni, M ;
Testoni, M ;
Toni, A .
BIOMATERIALS, 2001, 22 (07) :717-723
[5]
LONG-TERM RESULTS OF UNCEMENTED ALUMINA ACETABULAR IMPLANTS [J].
BOEHLER, M ;
KNAHR, K ;
PLENK, H ;
WALTER, A ;
SALZER, M ;
SCHREIBER, V .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :53-59
[6]
THE USE OF DENSE ALUMINA ALUMINA CERAMIC COMBINATION IN TOTAL HIP-REPLACEMENT [J].
BOUTIN, P ;
CHRISTEL, P ;
DORLOT, JM ;
MEUNIER, A ;
DEROQUANCOURT, A ;
BLANQUAERT, D ;
HERMAN, S ;
SEDEL, L ;
WITVOET, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12) :1203-1232
[7]
Uncertainty evaluation of volumetric wear assessment from coordinate measurements of ceramic hip joint prostheses [J].
Carmignato, S. ;
Spinelli, M. ;
Affatato, S. ;
Savio, E. .
WEAR, 2011, 270 (9-10) :584-590
[8]
Alumina-on-alumina total hip arthroplasty - A minimum 18.5-year follow-up study [J].
Hamadouche, M ;
Boutin, P ;
Daussange, J ;
Bolander, ME ;
Sedel, L .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (01) :69-77
[9]
*ISO, 1036042000 ISO
[10]
Macchi F., 2001, SCALPELLO, V15, P99