Is Ceramic-On-Ceramic Squeaking Phenomenon Reproducible In Vitro? A Long-Term Simulator Study Under Severe Conditions

被引:22
作者
Affatato, S. [1 ]
Traina, F. [1 ,2 ]
Mazzega-Fabbro, C. [1 ,3 ]
Sergo, V. [3 ]
Viceconti, M. [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, Bologna, Italy
[2] Prima Div & Ortopedia & Traumatol, Bologna, Italy
[3] Univ Trieste, DMRN, Trieste, Italy
关键词
hip implant; dislocation; inclination; ceramic wear; TOTAL HIP-ARTHROPLASTY; UHMWPE ACETABULAR CUPS; JOINT SIMULATOR; BEARING COUPLES; WEAR DEBRIS; PART; ALUMINA; POLYETHYLENE; REPLACEMENT; ORIENTATION;
D O I
10.1002/jbm.b.31398
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Clinical and in vitro studies on ceramic hip prostheses correlate cup implant position with hip noise, ceramic wear, or ceramic liner damage. A ceramic cup malposition could lead to edge load, ceramic head wear, and squeaking. A noise of a ceramic hip could also be correlate with implant instability and liner damage. Aim of this study was to investigate the long-term wear behavior of 12 commercial alumina-on-alumina bearings under severe conditions: different angles of inclination (23 degrees, 45 degrees, and 63 degrees) and the addition of third body particles (titanium and alumina powder) to address the effective role of cup position and ceramic particles on wear and hip noise. The study was performed using a 12-stations hip joint wear simulator (Shore Western, Monrovia) under bovine calf serum used as lubricant. Wear was evaluated by gravimetric method and the piezo-spectroscopic technique was used to evaluate the residual stress of the ceramic components and correlate this to the weight loss. After eight million cycles, we found that the inclination of the cup (63 degrees in this study) was the most disadvantaged and it was correlated with a hip noise. Gravimetric measurements showed higher wear than the other configurations and these results were in agreement with the Photoluminescence investigation. In particular, the results obtained in this work revealed a residual stress state greater not only with respect to the other angles of inclination but also to two retrieved alumina acetabular cups with a 10 years follow-up. (C) 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Bioniater 91B: 264-271, 2009
引用
收藏
页码:264 / 271
页数:8
相关论文
共 53 条
[1]
The predictive power of surface profile parameters on the amount of wear measured in vitro on metal-on-polyethylene artificial hip joints [J].
Affatato, S ;
Bersaglia, G ;
Yin, JQ ;
Traina, F ;
Toni, A ;
Viceconti, M .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H3) :457-464
[2]
Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions [J].
Affatato, S ;
Bersaglia, G ;
Rocchi, M ;
Taddei, P ;
Fagnano, C ;
Toni, A .
BIOMATERIALS, 2005, 26 (16) :3259-3267
[3]
The influence of implant position on the wear of alumina-on-alumina studied in a hip simulator [J].
Affatato, S ;
Bersaglia, G ;
Foltran, I ;
Emiliani, D ;
Traina, F ;
Toni, A .
WEAR, 2004, 256 (3-4) :400-405
[4]
The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2: wear particle characteristics with isolation protocols [J].
Affatato, S ;
Bersaglia, G ;
Emiliani, D ;
Foltran, I ;
Taddei, P ;
Reggiani, M ;
Ferrieri, P ;
Toni, A .
BIOMATERIALS, 2003, 24 (22) :4045-4055
[5]
The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: role of the third-body wear process [J].
Affatato, S ;
Bersaglia, G ;
Foltran, I ;
Taddei, P ;
Fini, G ;
Toni, A .
BIOMATERIALS, 2002, 23 (24) :4839-4846
[6]
Mixed-oxides prosthetic ceramic ball heads.: Part 2:: Effect of the ZrO2 fraction on the wear of ceramic on ceramic joints [J].
Affatato, S ;
Testoni, M ;
Cacciari, GL ;
Toni, A .
BIOMATERIALS, 1999, 20 (20) :1925-1929
[7]
The significance of metal staining on alumina femoral heads in total hip arthroplasty [J].
Bal, B. Sonny ;
Rahaman, Mohamed N. ;
Aleto, Thomas ;
Miller, F. Scott ;
Traina, Francesco ;
Toni, Aldo .
JOURNAL OF ARTHROPLASTY, 2007, 22 (01) :14-19
[8]
Concerns about ceramics in THA [J].
Barrack, RL ;
Barak, C ;
Skinner, HB .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :73-79
[9]
Wear debris from two different alumina-on-alumina total hip arthroplasties [J].
Böhler, M ;
Mochida, Y ;
Bauer, TW ;
Plenk, H ;
Salzer, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (06) :901-909
[10]
CHRISTEL PS, 1992, CLIN ORTHOP RELAT R, P10