Measurement of soluble L-lactate in dairy products using screen-printed sensors in batch mode

被引:26
作者
Collier, WA [1 ]
Janssen, D [1 ]
Hart, AL [1 ]
机构
[1] AGRES GRASSLANDS,PALMERSTON NORTH,NEW ZEALAND
关键词
lactate; lactate oxidase; screen-printing; dairy products;
D O I
10.1016/0956-5663(96)87663-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Amperometric sensors based on lactate oxidase and platinized carbon were constructed entirely by screen-printing. They were used to estimate the lactate concentration in diluted samples of yoghurt and buttermilk. Estimates made in untreated (apart from dilution) solutions were subject to bias, relative to values obtained from spectrophotometric assays. Removal of cations improved the accuracy of the estimates. Calcium may have made a small contribution to the bias but the nature of the inhibition is otherwise unknown. With untreated samples, determinations of any accuracy require sufficient dilution to avoid such bias, although the smaller signals resulting from dilution place more demands on calibration and manufacturing standards. The relative standard deviations of estimates, based on variation among sensors, of the most dilute samples were generally too high for practical use. The errors of estimates at lower dilutions approached those of screen-printed sensors made under industrial conditions. (C) 1996 Elsevier Science Limited.
引用
收藏
页码:1041 / 1049
页数:9
相关论文
共 25 条