A Miniature-Implantable RF-Wireless Active Glaucoma Intraocular Pressure Monitor

被引:112
作者
Chow, Eric Y. [1 ]
Chlebowski, Arthur L. [1 ]
Irazoqui, Pedro P. [1 ]
机构
[1] Purdue Univ, Ctr Implantable Devices, Weldon Sch Biomed Engn, W Lafayette, IN 47907 USA
关键词
Biomedical applications of electromagnetic radiation; biomedical monitoring; biomedical telemetry; implantable biomedical devices; DESIGN; SYSTEM;
D O I
10.1109/TBCAS.2010.2081364
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Glaucoma is a detrimental disease that causes blindness in millions of people worldwide. There are numerous treatments to slow the condition but none are totally effective and all have significant side effects. Currently, a continuous monitoring device is not available, but its development may open up new avenues for treatment. This work focuses on the design and fabrication of an active glaucoma intraocular pressure (IOP) monitor that is fully wireless and implantable. Major benefits of an active IOP monitoring device include the potential to operate independently from an external device for extended periods of time and the possibility of developing a closed-loop monitoring and treatment system. The fully wireless operation is based off using gigahertz-frequency electromagnetic wave propagation, which allows for an orientation independent transfer of power and data over reasonable distances. Our system is comprised of a micro-electromechanical systems (MEMS) pressure sensor, a capacitive power storage array, an application-specific integrated circuit designed on the Texas Instruments (TI) 130 nm process, and a monopole antenna all assembled into a biocompatible liquid-crystal polymer-based tad-pole-shaped package.
引用
收藏
页码:340 / 349
页数:10
相关论文
共 35 条
[1]  
[Anonymous], 2005, C951 IEEE INT COMM E
[2]  
BACKLUND LRY, 1990, SENSOR ACTUAT A-PHYS, P58
[3]   Microfabricated Implantable Parylene-Based Wireless Passive Intraocular Pressure Sensors [J].
Chen, Po-Jui ;
Rodger, Damien C. ;
Saati, Saloomeh ;
Humayun, Mark S. ;
Tai, Yu-Chong .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2008, 17 (06) :1342-1351
[4]  
CHOUARD CH, 1995, B ACAD NAT MED PARIS, V179, P549
[5]   Implantable Wireless Telemetry Boards for In Vivo Transocular Transmission [J].
Chow, Eric Y. ;
Yang, Chin-Lung ;
Chlebowski, Arthur ;
Moon, Sungwook ;
Chappell, William J. ;
Irazoqui, Pedro P. .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2008, 56 (12) :3200-3208
[6]  
CHOW EY, 2010, INT SOL STAT CIRC C
[7]   A model for μ-power rectifier analysis and design [J].
Curty, JP ;
Joehl, N ;
Krummenacher, F ;
Dehollain, C ;
Declercq, MJ .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2005, 52 (12) :2771-2779
[8]  
DeHennis A, 2002, PROC IEEE MICR ELECT, P252, DOI 10.1109/MEMSYS.2002.984250
[9]   Embedded 5 V to 3.3 V voltage regulator for supplying digital IC's in 3.3 V CMOS technology [J].
den Besten, GW ;
Nauta, B .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1998, 33 (07) :956-962
[10]   Wireless intra-ocular pressure monitoring system integrated into an artificial lens [J].
Eggers, T ;
Draeger, J ;
Hille, K ;
Marschner, C ;
Stegmaier, P ;
Binder, J ;
Laur, R .