Extraction of 15 microcystins and nodularin using immunoaffinity columns

被引:39
作者
Aranda-Rodriguez, R [1 ]
Kubwabo, C [1 ]
Benoit, FM [1 ]
机构
[1] Hlth Canada, Hlth Environm & Consumer Safety Branch, Safe Environm Programme, Div Chem Res, Ottawa, ON K1A 0L2, Canada
关键词
microcystins; nodularin; liquid chromatography-electrospray ionization mass spectrometry; solid-phase extraction; immunoaffinity columns; Oasis (R) HLB; PERFORMANCE LIQUID-CHROMATOGRAPHY; PHOSPHATASE INHIBITION ASSAY; LINKED-IMMUNOSORBENT-ASSAY; BLUE-GREEN-ALGAE; HEPTAPEPTIDE MICROCYSTINS; CYANOBACTERIAL TOXINS; MONOCLONAL-ANTIBODIES; MASS-SPECTROMETRY; LAKE WATER; CLEANUP;
D O I
10.1016/j.toxicon.2003.08.001
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Microcystins (MCYSTs) were isolated from surface water using reusable immunoaffinity columns. Individual MCYST were determined by high performance liquid chromatography equipped with a photo-diode array detector (HPLC-PDA, 200-300 nm). Subsequent analysis of the samples by liquid chromatography-electrospray ionization mass spectrometry (LC-ESMS) provided molecular weight information, which was used to tentatively identify individual MCYST variants for which standards were not available. Results obtained using immunoaffinity columns (IAC)-HPLC-PDA were compared to those obtained using solid phase extraction (SPE) Oasis(R) HLB-HPLC-PDA. This is the first report of the extraction of 15 microcystins and nodularin using immunoaffinity columns. Whereas previous reports demonstrates the use of IAC for four microcystins, we found that IAC selectively extracted the following microcystins: MCYST-RR, [D-Asp(3)]MCYST-RR, MCYST-YR, MCYST-LR, 3 MCYST-LR variants, MCYST-AR, MCYST-FR, MCYST-WR, MCYST-LA, MCYST-LA variant, the less polar microcystins such as MCYST-LF, MCYST-LW and nodularin. The IAC extracts were free of interferences which enabled better detection and identification of MCYSTs. Based on the amount loaded to the cartridges, the method detection limit was 10-14 ng when using IAC and 25 ng for SPE of each MCYST-RR, MCYST-YR and MCYST-LR. Reproducibilities expressed as relative standard deviation were 6-10% for SPE and 4-17% for IAC. Crown Copyright (C) 2003 Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:587 / 599
页数:13
相关论文
共 35 条
[21]  
2, 10.1002/etc.5620200805]
[22]   Detection of microcystins with protein phosphatase inhibition assay, high-performance liquid chromatography-UV detection and enzyme-linked immunosorbent assay - Comparison of methods [J].
Rapala, J ;
Erkomaa, K ;
Kukkonen, J ;
Sivonen, K ;
Lahti, K .
ANALYTICA CHIMICA ACTA, 2002, 466 (02) :213-231
[23]   Potential of immunoextraction coupled to analytical and bioanalytical methods (liquid chromatography, ELISA kit and phosphatase inhibition test) for an improved environmental monitoring of cyanobacterial toxins [J].
Rivasseau, C ;
Hennion, MC .
ANALYTICA CHIMICA ACTA, 1999, 399 (1-2) :75-87
[24]  
Sherlock Ian R., 1997, Natural Toxins, V5, P247, DOI 10.1002/(SICI)1522-7189(1997)5:6<247::AID-NT5>3.0.CO
[25]  
2-N
[26]   Separation, purity testing and identification of cyanobacterial hepatotoxins with capillary electrophoresis and electrospray mass spectrometry [J].
Sirén, H ;
Jussila, M ;
Liu, HW ;
Peltoniemi, S ;
Sivonen, K ;
Riekkola, ML .
JOURNAL OF CHROMATOGRAPHY A, 1999, 839 (1-2) :203-215
[27]   High-performance liquid chromatographic separation of microcystins and nodularin, cyanobacterial peptide toxins, on C18 and amide C16 sorbents [J].
Spoof, L ;
Karlsson, K ;
Meriluoto, J .
JOURNAL OF CHROMATOGRAPHY A, 2001, 909 (02) :225-236
[28]  
Tsuchiya Y, 1997, JPN J TOX ENV HEALTH, V43, P190
[29]   Immunoaffinity column as clean-up tool for determination of trace amounts of microcystins in tap water [J].
Tsutsumi, T ;
Nagata, S ;
Hasegawa, A ;
Ueno, Y .
FOOD AND CHEMICAL TOXICOLOGY, 2000, 38 (07) :593-597
[30]   Anti-idiotype monoclonal antibodies against anti-microcystin antibody and their use in enzyme immunoassay [J].
Tsutsumi, T ;
Nagata, S ;
Yoshida, F ;
Ueno, Y .
TOXICON, 1998, 36 (02) :235-245