Dual-labeled glucose binding protein for ratiometric measurements of glucose

被引:72
作者
Ge, XD [1 ]
Tolosa, L [1 ]
Rao, G [1 ]
机构
[1] Univ Maryland, Dept Chem & Biochem Engn, Baltimore, MD 21250 USA
关键词
D O I
10.1021/ac035063p
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Highly sensitive glucose monitoring has potential applications in conditions where the glucose levels are below the detection limit of currently available technology. Examples include bioprocess monitoring of bacterial cultures and measurement of minute amounts of human interstitial fluid extracted by iontophoresis. Here we describe a ratiometric glucose sensor capable of measuring micromolar levels of glucose. This sensor is based on an E. coli glucose binding protein (GBP) labeled with two fluoro-phores. The L255C mutant of GBP was labeled with the environment-sensitive fluorophore, acrylodan, at the cysteine mutation and a long-lived metal ligand complex of ruthenium (ruthenium bis(2,2'-bipyridyl)-1, 10-phenanthroline-9-isothiocyanate) at the N-terminal. The acrylodan emission is quenched in the presence of glucose while the ruthenium emission remained constant, thereby serving as a reference. The sensitivity of the sensor is in the micromolar range (K-d = 0.4-1.4 muM). Thus, it is possible to measure glucose concentrations at micromolar levels and higher (with dilution). Calculations of the fluorescence energy-transfer efficiency between acrylodan and ruthenium gave an approximate distance of 25 Angstrom between the two fluorophores, consistent with X-ray crystallographic data. The effect of temperature on glucose binding was measured and analyzed. Maximum signal changes and apparent binding constants increase with temperature. The enthalpy change for glucose binding as calculated from the apparent binding constants is similar to43.1 kJ/mol. In addition to ratiometric measurements, the presence of the long-lived ruthenium metal ligand complex allows for low-cost modulation-based sensing.
引用
收藏
页码:1403 / 1410
页数:8
相关论文
共 36 条
  • [1] A fluorescence affinity hollow fiber sensor for continuous transdermal glucose monitoring
    Ballerstadt, R
    Schultz, JS
    [J]. ANALYTICAL CHEMISTRY, 2000, 72 (17) : 4185 - 4192
  • [2] BOOS W, 1971, J BIOL CHEM, V246, P621
  • [3] In situ assembled mass-transport controlling micromembranes and their application in implanted amperometric glucose sensors
    Chen, T
    Friedman, KA
    Lei, I
    Heller, A
    [J]. ANALYTICAL CHEMISTRY, 2000, 72 (16) : 3757 - 3763
  • [4] Application of amperometric sol-gel biosensor to flow injection determination of glucose
    Couto, CMCM
    Araújo, AN
    Montenegro, MCBSM
    Rohwedder, J
    Raimundo, I
    Pasquini, C
    [J]. TALANTA, 2002, 56 (06) : 997 - 1003
  • [5] Optical determination of glutamine using a genetically engineered protein
    Dattelbaum, JD
    Lakowicz, JR
    [J]. ANALYTICAL BIOCHEMISTRY, 2001, 291 (01) : 89 - 95
  • [6] SCALE-UP OF THE ADENOVIRUS EXPRESSION SYSTEM FOR THE PRODUCTION OF RECOMBINANT PROTEIN IN HUMAN 293S CELLS
    GARNIER, A
    COTE, J
    NADEAU, I
    KAMEN, A
    MASSIE, B
    [J]. CYTOTECHNOLOGY, 1994, 15 (1-3) : 145 - 155
  • [7] Genetically engineered binding proteins as biosensors for fermentation and cell culture
    Ge, XD
    Tolosa, L
    Simpson, J
    Rao, G
    [J]. BIOTECHNOLOGY AND BIOENGINEERING, 2003, 84 (06) : 723 - 731
  • [8] Enhancement of operational stability of an enzyme biosensor for glucose and sucrose using protein based stabilizing agents
    Gouda, MD
    Kumar, MA
    Thakur, MS
    Karanth, NG
    [J]. BIOSENSORS & BIOELECTRONICS, 2002, 17 (6-7) : 503 - 507
  • [9] Continuous monitoring of D-glucose and L-lactate by flow injection analysis
    Haouz, A
    Stieg, S
    [J]. ENZYME AND MICROBIAL TECHNOLOGY, 2002, 30 (01) : 129 - 133
  • [10] Haughland R.P., 1996, HDB FLUORESCENT PROB