MR evaluation of ventricular function: True fast imaging with steady-state precession versus fast low-angle shot cine MR imaging: Feasibility study

被引:284
作者
Barkhausen, J
Ruehm, SG
Goyen, M
Buck, T
Laub, G
Debatin, JF
机构
[1] Univ Hosp Essen, Dept Diagnost Radiol, D-45122 Essen, Germany
[2] Univ Hosp Essen, Dept Cardiol, D-45122 Essen, Germany
[3] Siemens Med Syst, Erlangen, Germany
关键词
heart; MR; ventricles; volume; magnetic resonance (MR); cine study; pulse sequences; volume measurement;
D O I
10.1148/radiology.219.1.r01ap12264
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Short- and long-axis cine magnetic resonance (MR) images were obtained with a standard fast low-angle shot, or FLASH, sequence and a first generation true fast imaging with steady-state precession (FISP) sequence on a 1.5-T MR imager. Contrast-to-noise ratios and volumetric left ventricular measurements were compared for manual and automatic segmentation. True FISP images were associated with significantly (P <.01) higher contrast-to-noise ratios and allowed better detection of the endocardial border. True FISP images were provided with short acquisition times and excellent contrast between the myocardium and the ventricular lumen.
引用
收藏
页码:264 / 269
页数:6
相关论文
共 20 条
[11]   IMPROVED EJECTION FRACTION AND FLOW VELOCITY ESTIMATES WITH USE OF VIEW SHARING AND UNIFORM REPETITION TIME EXCITATION WITH FAST CARDIAC TECHNIQUES [J].
FOO, TKF ;
BERNSTEIN, MA ;
AISEN, AM ;
HERNANDEZ, RJ ;
COLLICK, BD ;
BERNSTEIN, T .
RADIOLOGY, 1995, 195 (02) :471-478
[12]   Experimental validation of an automated edge-detection method for a simultaneous determination of the endocardial and epicardial borders in short-axis cardiac MR images:: Application in normal volunteers [J].
Furber, A ;
Balzer, P ;
Cavaro-Ménard, C ;
Croué, A ;
Da Costa, E ;
Lethimonnier, F ;
Geslin, P ;
Tadéi, A ;
Jallet, P ;
Le Jeune, JJ .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (05) :1006-1014
[13]  
Heusch A, 1999, Eur J Ultrasound, V9, P245, DOI 10.1016/S0929-8266(99)00032-4
[14]   WHICH STANDARD HAS THE GOLD [J].
HIGGINS, CB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (07) :1608-1609
[15]   Automatic detection of left ventricular contours from cardiac cine magnetic resonance imaging using fuzzy logic [J].
Lalande, A ;
Legrand, L ;
Walker, PM ;
Guy, F ;
Cottin, Y ;
Roy, S ;
Brunotte, F .
INVESTIGATIVE RADIOLOGY, 1999, 34 (03) :211-217
[16]  
Rominger MB, 1999, ROFO-FORTSCHR RONTG, V170, P54
[17]   Automatic cardiac MR image segmentation using edge detection by tissue classification in pixel neighborhoods [J].
Singleton, HR ;
Pohost, GM .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (03) :418-424
[18]   Use of an intravascular T1 contrast agent to improve MR cine myocardial-blood pool definition in men [J].
Stillman, AE ;
Wilke, N ;
JeroschHerold, M .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (04) :765-767
[19]   CINE MR DETERMINATION OF LEFT-VENTRICULAR EJECTION FRACTION [J].
UTZ, JA ;
HERFKENS, RJ ;
HEINSIMER, JA ;
BASHORE, T ;
CALIFF, R ;
GLOVER, G ;
PELC, N ;
SHIMAKAWA, A .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 148 (05) :839-843
[20]   Comparison between manual and semiautomated analysis of left ventricular volume parameters from short-axis MR images [J].
vanderGeest, RJ ;
Buller, VGM ;
Jansen, E ;
Lamb, HJ ;
Baur, LHB ;
vanderWall, EE ;
deRoos, A ;
Reiber, JHC .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1997, 21 (05) :756-765