Microfluidic device for immunoassays based on surface plasmon resonance imaging

被引:121
作者
Luo, Yiqi [1 ]
Yu, Fang [1 ]
Zare, Richard N. [1 ]
机构
[1] Stanford Univ, Dept Chem, Stanford, CA 94305 USA
关键词
D O I
10.1039/b800606g
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have designed and fabricated a polydimethylsiloxane (PDMS) microfluidic device containing an array of gold spots onto which antigens or antibodies of interest can be attached. We use surface plasmon resonance (SPR) imaging to monitor the antibody-antigen recognition and binding events. This combination offers two significant advantages: (1) the microfluidic device dramatically reduces reaction time and sample consumption; and (2) the SPR imaging yields real-time detection of the immunocomplex formation. Thus, an immunoreaction may be detected and quantitatively characterized in about 10 min. The sensitivity of this method is at the subnanomolar level. When gold nanoparticles are selectively coupled to the immunocomplex to cause signal amplification, the sensitivity reaches the ten to one hundred picomolar level but the time required increases to about 60 min.
引用
收藏
页码:694 / 700
页数:7
相关论文
共 28 条
[1]   Studies on surface wettability of poly(dimethyl) siloxane (PDMS) and glass under oxygen-plasma treatment and correlation with bond strength [J].
Bhattacharya, S ;
Datta, A ;
Berg, JM ;
Gangopadhyay, S .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2005, 14 (03) :590-597
[2]   Exploring "one-shot" kinetics and small molecule analysis using the ProteOn XPR36 array biosensor [J].
Bravman, Tsafrir ;
Bronner, Vered ;
Lavie, Kobi ;
Notcovich, Ariel ;
Papalia, Giuseppe A. ;
Myszka, David G. .
ANALYTICAL BIOCHEMISTRY, 2006, 358 (02) :281-288
[3]  
Chiem NH, 1998, CLIN CHEM, V44, P591
[4]   Development and characterization of an ELISA assay in PDMS microfluidic channels [J].
Eteshola, E ;
Leckband, D .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 72 (02) :129-133
[5]   Flow injection immunoassays: a review [J].
Fintschenko, Y ;
Wilson, GS .
MIKROCHIMICA ACTA, 1998, 129 (1-2) :7-18
[6]   SPR imaging-based salivary diagnostics system for the detection of small molecule analytes [J].
Fu, Elain ;
Chinowsky, Timothy ;
Nelson, Kjell ;
Johnston, Kyle ;
Edwards, Thayne ;
Helton, Kristen ;
Grow, Michael ;
Miller, John W. ;
Yager, Paul .
ORAL-BASED DIAGNOSTICS, 2007, 1098 :335-344
[7]   Colloidal Au-enhanced surface plasmon resonance for ultrasensitive detection of DNA hybridization [J].
He, L ;
Musick, MD ;
Nicewarner, SR ;
Salinas, FG ;
Benkovic, SJ ;
Natan, MJ ;
Keating, CD .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2000, 122 (38) :9071-9077
[8]   Coating of poly(dimethylsiloxane) with n-dodecyl-β-D-maltoside to minimize nonspecific protein adsorption [J].
Huang, B ;
Wu, HK ;
Kim, S ;
Zare, RN .
LAB ON A CHIP, 2005, 5 (10) :1005-1007
[9]   Sensitivity enhancement of optical immunosensors with nanoparticles [J].
Kubitschko, S ;
Spinke, J ;
Bruckner, T ;
Pohl, S ;
Oranth, N .
ANALYTICAL BIOCHEMISTRY, 1997, 253 (01) :112-122
[10]   On-chip enzyme immunoassay of a cardiac marker using a microfluidic device combined with a portable surface plasmon resonance system [J].
Kurita, Ryoji ;
Yokota, Yoshimi ;
Sato, Yukari ;
Mizutani, Fumio ;
Niwa, Osamu .
ANALYTICAL CHEMISTRY, 2006, 78 (15) :5525-5531