Renal cyst pseudoenhancement: Influence of multidetector CT reconstruction algorithm and scanner type in phantom model

被引:72
作者
Birnbaum, Bernard A. [1 ]
Hindman, Nicole [1 ]
Lee, Julie [1 ]
Babb, James S. [1 ]
机构
[1] NYU Med Ctr, Dept Radiol, New York, NY 10016 USA
关键词
D O I
10.1148/radiol.2443061537
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prospectively determine the dependence of renal cyst pseudoenhancement on multidetector computed tomographic (CT) scanner type and convolution kernel in a phantom model. Materials and Methods: A customized anthropomorphic phantom was created to accept interchangeable 40-, 140-, and 240-HU renal inserts that contained stacked 0- and 50-HU cylindric cysts measuring 7, 10, and 15 mm in diameter. Each phantom and insert was scanned with five different multidetector CT scanners on five separate occasions by using 120 kVp, low and high tube current settings, 3.00-3.75-mm collimation, and standard and high-spatial-resolution kernels. A total of 2340 CT attenuation measurements were obtained by using standardized regions of interest. The effect of multidetector CT imaging regimen, tube current, cyst diameter, and renal attenuation on pseudoenhancement incidence was assessed by using generalized estimating equations based on a binary logistic regression model. Within this framework, a Bonferroni multiple comparison correction was used to assess pseudoenhancement frequency differences among imaging regimens. Results: Pseudoenhancement occurred in both 0- and 50-HU cysts; was significantly correlated with multidetector CT imaging regimen (P < .0001), cyst diameter (P < .0001), and renal attenuation (P <= .032); and was independent of tube current (P > .3). When convolution kernels on specific scanners were compared, significant differences (P < .04) between kernels were identified with all five scanners in terms of. observed pseudoenhancement incidence. Generational differences in equipment were noted, with pseudoenhancement incidence ranging from 1.7% to 8.3%, 1.7% to 16.7%, and 18.3% to 56.7% across relevant kernels for three scanners from one manufacturer. Conclusion: Pseudoenhancement is strongly dependent on multidetector CT convolution kernel. Varying this parameter may mitigate this phenomenon, which is independent of volume-averaging effects.
引用
收藏
页码:767 / 775
页数:9
相关论文
共 9 条
[1]   Pseudoenhancement of simulated renal cysts in a phantom using different multidetector CT scanners [J].
Abdulla, C ;
Kalra, MK ;
Saini, S ;
Maher, MM ;
Ahmad, A ;
Halpern, E ;
Silverman, SG .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (06) :1473-1476
[2]   Renal cysts: Is attenuation artifactually increased on contrast-enhanced CT images? [J].
Bae, KT ;
Heiken, JP ;
Siegel, CL ;
Bennett, HF .
RADIOLOGY, 2000, 216 (03) :792-796
[3]   Renal cyst pseudoenhancement: Evaluation with an anthropomorphic body CT phantom [J].
Birnbaum, BA ;
Maki, DD ;
Chakraborty, DP ;
Jacobs, JE ;
Babb, JS .
RADIOLOGY, 2002, 225 (01) :83-90
[4]  
BOSNIAK MA, 1986, RADIOLOGY, V158, P1
[6]   Evaluation of pseudoenhancement of renal cysts during contrast-enhanced CT [J].
Coulam, CH ;
Sheafor, DH ;
Leder, RA ;
Paulson, EK ;
DeLong, DM ;
Nelson, RC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) :493-498
[7]  
Gokan Takehiko, 2002, Radiat Med, V20, P187
[8]   Pseudoenhancement of simple renal cysts: A comparison of single and multidetector helical CT [J].
Heneghan, JP ;
Spielmann, AL ;
Sheafor, DH ;
Kliewer, MA ;
DeLong, DM ;
Nelson, RC .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2002, 26 (01) :90-94
[9]   Renal cyst pseudoenhancement: Beam-hardening effects on CT numbers [J].
Maki, DD ;
Birnbaum, BA ;
Chakraborty, DP ;
Jacobs, JE ;
Carvalho, BM ;
Herman, GT .
RADIOLOGY, 1999, 213 (02) :468-472