Wear of congruent meniscal bearings in unicompartmental knee arthroplasty - A retrieval study of 16 specimens

被引:146
作者
Psychoyios, V [1 ]
Crawford, RW [1 ]
O'Connor, JJ [1 ]
Murray, DW [1 ]
机构
[1] Nuffield Orthopaed Ctr, Oxford Orthopaed Engn Ctr, Oxford OX3 7LD, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1998年 / 80B卷 / 06期
关键词
D O I
10.1302/0301-620X.80B6.8176
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements, We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm psr year, In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear.
引用
收藏
页码:976 / 982
页数:7
相关论文
共 17 条
[1]   POLYETHYLENE WEAR IN MENISCAL KNEE REPLACEMENT - A ONE TO 9-YEAR RETRIEVAL ANALYSIS OF THE OXFORD KNEE [J].
ARGENSON, JN ;
OCONNOR, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (02) :228-232
[2]  
BARTLEY RE, 1994, CLIN ORTHOP RELAT R, V299, P18
[3]   POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED [J].
BLUNN, GW ;
JOSHI, AB ;
LILLEY, PA ;
ENGELBRECHT, E ;
RYD, L ;
LIDGREN, L ;
HARDINGE, K ;
NIEDER, E ;
WALKER, PS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03) :247-255
[4]  
Cameron H U, 1988, Orthop Rev, V17, P983
[5]   POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[6]  
HALL RM, 1998, J BONE JOINT SU B S1, V80, P39
[7]  
HECK DA, 1993, CLIN ORTHOP RELAT R, V286, P154
[8]   INVIVO WEAR OF POLYETHYLENE ACETABULAR COMPONENTS [J].
KABO, JM ;
GEBHARD, JS ;
LOREN, G ;
AMSTUTZ, HC .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02) :254-258
[9]   THE SWEDISH KNEE ARTHROPLASTY REGISTER - A NATIONWIDE STUDY OF 30,003 KNEES 1976-1992 [J].
KNUTSON, K ;
LEWOLD, S ;
ROBERTSSON, O ;
LIDGREN, L .
ACTA ORTHOPAEDICA SCANDINAVICA, 1994, 65 (04) :375-386
[10]   Oxford meniscal bearing knee versus the Marmor knee in unicompartmental arthroplasty for arthrosis - A Swedish multicenter survival study [J].
Lewold, S ;
Goodman, S ;
Knutson, K ;
Robertsson, O ;
Lidgren, L .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :722-731