Lab-scale fermentation tests of microchip with integrated electrochemical sensors for pH, temperature, dissolved oxygen and viable biomass concentration

被引:50
作者
Krommenhoek, Erik E. [1 ]
van Lceuwen, Michiel [2 ]
Gardeniers, Han [1 ]
van Gulik, Walter M. [2 ]
van den Berg, Albert [1 ]
Li, Xiaonan [2 ]
Ottens, Marcel [2 ]
van der Wielen, Luuk A. M. [2 ]
Heiinen, Joseph J. [2 ]
机构
[1] Univ Twente, MESA Inst Nenotechnol, NL-7500 AE Enschede, Netherlands
[2] Delft Univ Technol, Dept Biotechnol, Delft, Netherlands
关键词
batch fermentation; integrated electrochemical sensors; ultra-microelectrode array; impedance spectroscopy;
D O I
10.1002/bit.21661
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
This article shows the development and testing of a microchip with integrated electrochemical sensors for measurement of pH, temperature, dissolved oxygen and viable biomass concentration under yeast cultivation conditions. Measurements were done both under dynamic batch conditions as well as under prolonged continuous cultivation conditions. The response of the sensors compared well with conventional measurement techniques. The biomass sensor was based on impedance spectroscopy. The results of the biomass sensor matched very well with dry weight measurements and showed a limit of detection of similar to 1 g/L. The dissolved oxygen concentration was monitored amperometrically using an ultra-microelectrode array, which showed an accuracy of similar to 0.2 mg/L and negligible drift. pH was monitored using an ISFET with an accuracy well below 0.1 pH unit. The platinum thin-film temperature resistor followed temperature changes with similar to 0.1 degrees C accuracy. The dimensions of the multi sensor chip are chosen as such that it is compatible with the 96-well plate format.
引用
收藏
页码:884 / 892
页数:9
相关论文
共 39 条
[21]   Minibioreactors [J].
Kumar, S ;
Wittmann, C ;
Heinzle, E .
BIOTECHNOLOGY LETTERS, 2004, 26 (01) :1-10
[22]   Design of a prototype miniature bioreactor for high throughput automated bioprocessing [J].
Lamping, SR ;
Zhang, H ;
Allen, B ;
Shamlou, PA .
CHEMICAL ENGINEERING SCIENCE, 2003, 58 (3-6) :747-758
[23]   Microbioreactor arrays with integrated mixers and fluid injectors for high-throughput experimentation with pH and dissolved oxygen control [J].
Lee, Harry L. T. ;
Boccazzi, Paolo ;
Ram, Rajeev J. ;
Sinskey, Anthony J. .
LAB ON A CHIP, 2006, 6 (09) :1229-1235
[24]   Accelerated design of bioconversion processes using automated microscale processing techniques [J].
Lye, GJ ;
Ayazi-Shamlou, P ;
Baganz, F ;
Dalby, PA ;
Woodley, JM .
TRENDS IN BIOTECHNOLOGY, 2003, 21 (01) :29-37
[25]  
Maharbiz MM, 2004, BIOTECHNOL BIOENG, V85, P376, DOI 10.1002/bit.10835
[27]   Performance of amperometric sensors based on multiple microelectrode arrays [J].
Morf, WE ;
de Rooij, NF .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 44 (1-3) :538-541
[28]   NEW OXYGEN SENSORS AND THEIR APPLICATION TO BIOSENSING [J].
PAPKOVSKY, DB .
SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) :213-218
[29]  
RUBINSTEIN I, 1995, PHYS ELECTROCHEMISTR, P134
[30]   High-throughput phenomics: experimental methods for mapping fluxomes [J].
Sauer, U .
CURRENT OPINION IN BIOTECHNOLOGY, 2004, 15 (01) :58-63