Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode

被引:147
作者
LoSasso, T [1 ]
Chui, CS [1 ]
Ling, CC [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
关键词
multileaf collimation; intensity modulation; dosimetry; quality assurance;
D O I
10.1118/1.1410123
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The delivery of intensity modulated radiation therapy (IMRT) using dynamic multileaf collimation (DMLC) accounts for about 40% of all treatments on eight multileaf collimator (MLC)-equipped linacs currently at the central and satellite facilities of the Memorial Sloan Kettering Cancer Center (MSKCC). These numbers have been steadily growing for the past 6 years. Treatment with IMRT has been used for a variety of cancers, including tumors of the prostate, head and neck, breast, paraspinal regions, and brain. To ensure that patients are treated in an accurate, efficient, and safe manner, we have developed a comprehensive QA program including mechanical quality assurance, which monitors known potential weaknesses in the MLC device itself, and verification of patient-specific treatments, to confirm the reliability of new software and to identify errors introduced through human interaction with individual patient's plans. This paper presents the reasoning, methodology, and results for mechanical, dosimetric, and electronic tests that are conducted at MSKCC. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:2209 / 2219
页数:11
相关论文
共 11 条
[1]   A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy [J].
Arnfield, MR ;
Siebers, JV ;
Kim, JO ;
Wu, QW ;
Keall, PJ ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (10) :2231-2241
[2]   Requirements for leaf position accuracy for dynamic multileaf collimation [J].
Budgell, GJ ;
Mott, JHL ;
Williams, PC ;
Brown, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1211-1227
[3]   Relative profile and dose verification of intensity-modulated radiation therapy [J].
Chang, JW ;
Mageras, GS ;
Chui, CS ;
Ling, CC ;
Lutz, W .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (01) :231-240
[4]   Testing of dynamic multileaf collimation [J].
Chui, CS ;
Spirou, S ;
LoSasso, T .
MEDICAL PHYSICS, 1996, 23 (05) :635-641
[5]   A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assurance [J].
Kung, JH ;
Chen, GTY ;
Kuchnir, FK .
MEDICAL PHYSICS, 2000, 27 (10) :2226-2230
[6]   Conformal radiation treatment of prostate cancer using inversely-planned intensity-modulated photon beams produced with dynamic multileaf collimation [J].
Ling, CC ;
Burman, C ;
Chui, CS ;
Kutcher, GJ ;
Leibel, SA ;
LoSasso, T ;
Mohan, R ;
Bortfeld, T ;
Reinstein, L ;
Spirou, S ;
Wang, XH ;
Wu, QW ;
Zelefsky, M ;
Fuks, Z .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (04) :721-730
[7]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[8]   Dosimetric verification of intensity-modulated fields [J].
Wang, XH ;
Spirou, S ;
LoSasso, T ;
Stein, J ;
Chui, CS ;
Mohan, R .
MEDICAL PHYSICS, 1996, 23 (03) :317-327
[9]   Monitor unit calculation for an intensity modulated photon held by a simple scatter-summation algorithm [J].
Xing, L ;
Chen, Y ;
Luxton, G ;
Li, JG ;
Boyer, AL .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03) :N1-N7
[10]  
YORKE E, 2000, MED PHYS, V27, P1419