fat-water separation;
Dixon method;
in-phase and out-of-phase images;
steady-state free precession;
frequency offset;
D O I:
10.1002/mrm.10686
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
In this work the feasibility of separating fat and water signals using the balanced steady-state free precession (SSFP) technique is demonstrated. The technique is based on the observation (Scheffler and Hennig, Magnetic Resonance in Medicine 2003;49:395-397) that at the nominal values of TE = TR/2 in SSFP imaging, phase coherence can be achieved at essentially only two orientations (0degrees and 180degrees) relative to the RF pulses in the rotating frame, under the assumption of TR << T-2, and independently of the SSFP angle. This property allows in-phase and out-of-phase SSFP images to be obtained by proper choices of the center frequency offset, and thus allows the Dixon subtraction method to be utilized for effective fat-water separation. The TR and frequency offset for optimal fat-water separation are derived from theories. Experimental results from healthy subjects, using a 3.0 Tesla system, show that nearly complete fat suppression can be accomplished. (C) 2004 Wiley-Liss, Inc.