Intensity modulated proton therapy and its sensitivity to treatment uncertainties 1: the potential effects of calculational uncertainties

被引:299
作者
Lomax, A. J. [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Radiotherapy, CH-5232 Villigen, Switzerland
关键词
D O I
10.1088/0031-9155/53/4/014
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effects of calculational uncertainties on 3D and distal edge tracking ( DET) intensity modulated proton therapy ( IMPT) treatment plans have been investigated. Dose calculation uncertainties have been assessed by comparing analytical and Monte Carlo dose calculations, and potential range uncertainties by recalculating plans with all CT values modified by +/- 3%. Analysis of the volume of PTV agreeing to within +/- 3% between the two calculations shows that the 3D approach provides significantly improved agreement ( 87.1 versus 80.3% of points for the 3D and DET approaches, respectively). For the DET approach, doses in the CTV have also been found to globally change by 5% as a result of 3% changes in CT value. When varying the intra- field gradients of the plans a similar trend is seen, but with the more complex plans also being found to be more sensitive to both uncertainties. In conclusion, the DET approach has been found to be relatively sensitive to the calculational errors investigated here. In contrast, the 3D approach appears to be quite robust, unless strong internal gradients are present. Nevertheless, the routine use of uncertainty analysis is advised when assessing all forms of IMPT plans.
引用
收藏
页码:1027 / 1042
页数:16
相关论文
共 42 条
[1]   Target margins for random geometrical treatment uncertainties in conformal radiotherapy [J].
Bel, A ;
vanHerk, M ;
Lebesque, JV .
MEDICAL PHYSICS, 1996, 23 (09) :1537-1545
[2]  
BOSSHARDT M, 2005, IMPTESTRO PHYS C LIS
[3]   The influence of beam model differences in the comparison of dose calculation algorithms for lung cancer treatment planning [J].
Chetty, IJ ;
Rosu, M ;
McShan, DL ;
Fraass, BA ;
Ten Haken, RK .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05) :801-815
[4]   Accounting for center-of-mass target motion using convolution methods in Monte Carlo-based dose calculations of the lung [J].
Chetty, IJ ;
Rosu, M ;
McShan, DL ;
Fraass, BA ;
Balter, JM ;
Ten Haken, RK .
MEDICAL PHYSICS, 2004, 31 (04) :925-932
[5]  
Deasy JO, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P406
[6]   INFLUENCE OF THICK INHOMOGENEITIES ON CHARGED-PARTICLE BEAMS [J].
GOITEIN, M ;
SISTERSON, JM .
RADIATION RESEARCH, 1978, 74 (02) :217-230
[7]  
GOITEIN M, 1978, Medical Physics (Woodbury), V5, P258, DOI 10.1118/1.594507
[8]  
GOITEIN M, 1978, Medical Physics (Woodbury), V5, P265, DOI 10.1118/1.594509
[9]   CALCULATION OF THE UNCERTAINTY IN THE DOSE DELIVERED DURING RADIATION-THERAPY [J].
GOITEIN, M .
MEDICAL PHYSICS, 1985, 12 (05) :608-612
[10]   A simulation of the effects of set-up error and changes in breast volume on conventional and intensity-modulated treatments in breast radiotherapy [J].
Hector, C ;
Webb, S ;
Evans, PM .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) :1451-1471