Determination of dopamine in synthetic cerebrospinal fluid by SWV with a graphite-polyurethane composite electrode

被引:54
作者
de Toledo, RA
Santos, MC
Cavalheiro, ETG
Mazo, LH
机构
[1] Inst Quim Sao Carlos, Dept Quim Fis, Grp Mat Electroquim & Metodos Electroanalit, BR-13560970 Sao Paulo, Brazil
[2] Univ Fed Sao Carlos, CMDMC, Dept Quim, Lab Interdisciplinar Electroquim & Ceram, BR-13565905 Sao Paulo, Brazil
[3] Inst Quim Sao Carlos, Dept Quim & Fis Mol, BR-13560970 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
dopamine; electrochemical oxidation; analytical determination; graphite-polyurethane composite electrode; synthetic cerebrospinal fluid;
D O I
10.1007/s00216-005-3066-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This work describes an electroanalytical investigation of dopamine using cyclic voltammetry ( CV) and the graphite-polyurethane composite electrode (GPU). In CV studies, well-defined redox peaks characterize the oxidation process at the GPU electrode, which is indicative of electrocatalytic effects associated with active sites on the GPU electrode surface. A new analytical methodology was developed using the GPU electrode and square wave voltammetry (SWV) in BR buffer solution (0.1 mol L-1; pH 7.4). Analytical curves were constructed under optimized conditions ( f= 60s(-1), Delta E-a= 50 mV, Delta E-I= 2 mV) and detection and quantification limits of 6.4x10(-8) mol L-1 (12.1 mu g L-1) and 5.2x10(-6) mol L-1 (0.9 mg L-1), respectively, were achieved. The precision of the method was checked by performing ten successive measurements for a 9.9x10(-6) mol L-1 dopamine solution. For intra-assay and interassay precisions, the relative standard deviations were 1.9 and 2.3%, respectively. In order to evaluate the developed methodology, the determination of dopamine was performed with good sensitivity and selectivity, without the interference of ascorbic acid in synthetic cerebrospinal. fluid, which indicates that the new methodology enables reliable analysis of dopamine.
引用
收藏
页码:1161 / 1166
页数:6
相关论文
共 23 条
[1]  
Bard A. J., 1980, ELECTROCHEMICAL METH
[2]   Effect of cocaine, nomifensine, GBR 12909 and WIN 35428 on carbon fiber microelectrode sensitivity for voltammetric recording of dopamine [J].
Davidson, C ;
Ellinwood, EH ;
Douglas, SB ;
Lee, TH .
JOURNAL OF NEUROSCIENCE METHODS, 2000, 101 (01) :75-83
[3]   Study of the composite electrodes carbon-polyvinyl chloride and carbon-polyvinyl chloride/Nafion by ex situ and in situ methods [J].
Dávila, MM ;
Elizalde, MP ;
Mattusch, J ;
Wennrich, R .
ELECTROCHIMICA ACTA, 2001, 46 (20-21) :3189-3197
[4]   Dopamine voltammetry at overoxidised polyindole electrodes [J].
Ghita, M ;
Arrigan, DWM .
ELECTROCHIMICA ACTA, 2004, 49 (26) :4743-4751
[5]  
Gosser D., 1994, CYCLIC VOLTAMMETRY
[6]   CHARACTERIZATION OF AMPEROMETRY FOR IN-VIVO MEASUREMENT OF DOPAMINE DYNAMICS IN THE RAT-BRAIN [J].
KAWAGOE, KT ;
WIGHTMAN, RM .
TALANTA, 1994, 41 (06) :865-874
[7]   Electroanalysis of dopamine at a gold electrode modified with N-acetylcysteine self-assembled monolayer [J].
Liu, T ;
Li, MX ;
Li, QY .
TALANTA, 2004, 63 (04) :1053-1059
[8]   SQUARE-WAVE VOLTAMMETRY OF AN ADSORBED REACTANT [J].
LOVRIC, M ;
KOMORSKYLOVRIC, S .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1988, 248 (02) :239-253
[9]   Electrochemical response of ascorbic acid at conducting and electrogenerated polymer modified electrodes for electroanalytical applications:: a review [J].
Malinauskas, A ;
Garjonyté, R ;
Mazeikiene, R ;
Jureviciute, I .
TALANTA, 2004, 64 (01) :121-129
[10]   Evaluation of a new rigid carbon-castor oil polyurethane composite as an electrode material [J].
Mendes, RK ;
Claro-Neto, S ;
Cavalheiro, ETG .
TALANTA, 2002, 57 (05) :909-917