Dealing with uncertainty in diffusion tensor MR data

被引:9
作者
Basser, PJ
Pajevic, S
机构
[1] NICHD, Sect Tissue Biophys & Biomimet, LIMB, NIH, Bethesda, MD 20892 USA
[2] CIT, Math & Stat Comp Lab, NIH, Bethesda, MD 20892 USA
关键词
D O I
10.1560/8P01-8RMA-B9EH-6F4W
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This paper explains how radio frequency (RF) background noise produces uncertainty in measured diffusion tensor MRI (DT-MRI) data, how this noise can be modeled, and how its effects can be mitigated. DT-MRI data are derived from a series of magnitude diffusion-weighted images (DWI) in which RF noise is rectified. A new Gaussian distribution is proposed that describes the variability of the estimated diffusion tensor, D, in an ideal experiment in which RF noise is the only artifact present. We show how to improve the design of DT-MRI experiments by requiring that the statistical distribution of D be independent of the laboratory coordinate system. Non-parametric empirical methods of analyzing uncertainty in DT-MRI experiments are also described. Monte Carlo simulations are useful in designing and interpreting DT-MRI experiments. Bootstrap methods help us measure the true variability of D (and quantities derived from it), and assess the quality of DT-MRI data. Matrix Perturbation techniques predict how the uncertainty in D propagates to its eigenvalues and eigenvectors. A method for obtaining a continuous diffusion tensor field from the measured discrete noisy DT-MRI data also reduces the uncertainty of D and quantities derived from it. Finally, we describe schemes that use wavelets to remove noise from DWI and DT-MRI data while preserving boundaries between different tissue regions. Collectively, these parametric and nonparametric methods provide a unified statistical framework to improve the design of DT-MRI experiments and their subsequent analysis.
引用
收藏
页码:129 / 144
页数:16
相关论文
共 144 条
  • [21] BASSER PJ, 1999, 7 ANN M ISMRM PHIL, P1789
  • [22] BASSER PJ, 1998, 6 ISMRM SYDN AU, P1226
  • [23] BASSER PJ, 1996, Patent No. 5539310
  • [24] BASSER PJ, 1992, 11 ANN M SMRM BERL, P1221
  • [25] Correction of eddy current-induced artefacts in diffusion tensor imaging using iterative cross-correlation
    Bastin, ME
    [J]. MAGNETIC RESONANCE IMAGING, 1999, 17 (07) : 1011 - 1024
  • [26] BEHRENS TEJ, 2002, 11 ANN M ISMRM HON, P1142
  • [27] BEHRENS TEJ, 2002, 11 ANN M ISMRM HON, P1160
  • [28] BLOCH F, 1946, PHYS REV, V70, P460, DOI 10.1103/PhysRev.70.460
  • [29] High-resolution diffusion imaging using phase-corrected segmented echo-planar imaging
    Brockstedt, S
    Moore, JR
    Thomsen, C
    Holtås, S
    Ståhlberg, F
    [J]. MAGNETIC RESONANCE IMAGING, 2000, 18 (06) : 649 - 657
  • [30] MRI white matter diffusion anisotropy and PET metabolic rate in schizophrenia
    Buchsbaum, MS
    Tang, CY
    Peled, S
    Gudbjartsson, H
    Lu, DF
    Hazlett, EA
    Downhill, J
    Haznedar, M
    Fallon, JH
    Atlas, SW
    [J]. NEUROREPORT, 1998, 9 (03) : 425 - 430