A sample introduction method based on negative pressure in flow injection-capillary electrophoresis system and its application to the alkaline-earth metal cation separation

被引:15
作者
Wang, JW
Cai, PX
Mo, JY
Chen, ZG [1 ]
机构
[1] Sun Yat Sen Univ, Sch Pharmaceut Sci, Guangzhou 510089, Peoples R China
[2] Sun Yat Sen Univ, Sch Chem & Chem Engn, Guangzhou 510089, Peoples R China
基金
中国国家自然科学基金;
关键词
capillary electrophoresis; flow injection; negative pressure; contactless conductivity detector;
D O I
10.1081/AL-200051062
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new sample introduction method based on negative pressure has been developed in the flow injection (FI) - capillary electrophoresis (CE) system with contactless conductivity detection (CCD). In the system, the sample was carried to the interface by background electrolyte (BGE), which was carried by a dual-channel peristaltic pump. Sample introduction into capillary was performed through the negative pressure at the outlet of capillary. Duration of the injection was controlled by a precise homemade time circuit connected to the voltage supply of vacuum pump. When the sample was transported to the inlet of capillary, the vacuum pump was started and a small volume sample was injected within 1.0 - 2.0 s. Negative pressures from 20 to 25 kPa are appropriate and other important injection parameters were studied. The RSD values ( n = 8) of the negatively pressurized injections, within the range of 0.8 - 1.1% for migration time, are comparable with those of gravity injection. The performance of the whole system was further demonstrated with satisfactory results in the separation of four alkaline - earth metal cations ( Mg2+, Ca2+, Sr2+, Ba2+) with CCD.
引用
收藏
页码:857 / 867
页数:11
相关论文
共 22 条
[11]   New interface for coupling flow-injection and capillary electrophoresis [J].
Kuban, P ;
Engstrom, A ;
Olsson, JC ;
Thorsen, G ;
Tryzell, R ;
Karlberg, B .
ANALYTICA CHIMICA ACTA, 1997, 337 (02) :117-124
[12]   On-line monitoring of kraft pulping liquors with a valveless flow injection-capillary electrophoresis system [J].
Kuban, P ;
Karlberg, B .
ANALYTICA CHIMICA ACTA, 2000, 404 (01) :19-28
[13]   Flow injection analysis capillary electrophoresis system with hydrodynamic injection [J].
Kuban, P ;
Pirmohammadi, R ;
Karlberg, B .
ANALYTICA CHIMICA ACTA, 1999, 378 (1-3) :55-62
[14]  
Kuldvee R, 2000, ELECTROPHORESIS, V21, P2879, DOI 10.1002/1522-2683(20000801)21:14<2879::AID-ELPS2879>3.0.CO
[15]  
2-B
[16]   Gradient elution electrochromatography with a flow-injection analysis interface [J].
Lister, AS ;
Rimmer, CA ;
Dorsey, JG .
JOURNAL OF CHROMATOGRAPHY A, 1998, 828 (1-2) :105-112
[17]   Combination of flow injection with capillary electrophoresis. Part 6. A bias-free sample introduction system based on electroosmotic-flow traction [J].
Pu, QS ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 398 (01) :65-74
[18]   ROTARY-TYPE INJECTOR FOR CAPILLARY ZONE ELECTROPHORESIS [J].
TSUDA, T ;
MIZUNO, T ;
AKIYAMA, J .
ANALYTICAL CHEMISTRY, 1987, 59 (05) :799-800
[19]   SPLIT INJECTOR FOR CAPILLARY ZONE ELECTROPHORESIS [J].
TSUDA, T ;
ZARE, RN .
JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2) :103-110
[20]   Mobility-based selective on-line preconcentration of proteins in capillary electrophoresis by controlling electroosmotic flow [J].
Wang, QG ;
Yue, BF ;
Lee, ML .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1025 (01) :139-146