Transmission scan truncation with small-field-of-view dedicated cardiac SPECT systems: Impact and automated quality control

被引:10
作者
Chen, J
Galt, JR
Case, JA
Ye, JH
Cullom, SJ
Durbin, MK
Shao, L
Garcia, EV
机构
[1] Emory Univ, Dept Radiol, Atlanta, GA 30322 USA
[2] Cardiovasc Imaging Technol, Kansas City, MO USA
[3] Philips Med Syst, Milpitas, CA USA
关键词
single photon emission computed tomography; small field-of-view systems; attenuation correction; automated transmission scan truncation quality control;
D O I
10.1016/j.nuclcard.2005.04.013
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Small-field-of-view (FOV) dedicated cardiac single photon emission computed tomography (SPECT) systems will frequently exhibit severe transmission scan truncation that may degrade attenuation correction (AC). This study evaluated the impact of transmission scan truncation on AC and developed automated transmission scan truncation quality control (ATSTQC) for small-FOV systems. Methods and Results. Small-FOV data were simulated from the data of 10 patients acquired by a full-FOV Philips Vertex system. AC images of the full- and small-FOV data were compared by mean and maximum absolute differences of myocardial counts, and differences in stress and rest severity scores were calculated by use of the Emory Cardiac Toolbox.small-FOV systems. ATSTQC was developed to identify critical truncation that significantly increased these indices and then tested with 18 independent patients. Left-side truncation resulted in significant distortion of the quantitative indices. ATSTQC, developed on the condition that left-side truncation is critical, showed high concordance with the qualitative assessment in identification of critical truncation. Conclusions. Identification of left-side truncation as critical truncation is necessary to judge whether accurate AC can be obtained. The developed ATSTQC can accurately detect critical truncation and will help clinicians decide whether to use AC in a particular study.
引用
收藏
页码:567 / 573
页数:7
相关论文
共 26 条
[21]  
LUI Y, 2003, J NUCL MED, V44
[22]   QUANTITATIVE-ANALYSIS OF IMAGE TRUNCATION IN FOCAL-BEAM CT [J].
MANGLOS, SH ;
GAGNE, GM ;
BASSANO, DA .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (10) :1443-1457
[23]   Coronary calcium screening in asymptomatic patients as a guide to risk factor modification and stress myocardial perfusion imaging [J].
Moser, KW ;
O'Keefe, JH ;
Bateman, TM ;
McGhie, IA .
JOURNAL OF NUCLEAR CARDIOLOGY, 2003, 10 (06) :590-598
[24]   A two-step Hilbert transform method for 2D image reconstruction [J].
Noo, F ;
Clackdoyle, R ;
Pack, JD .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17) :3903-3923
[25]   Improved coronary disease detection with quantitative attenuation-corrected Tl-201 images [J].
Shotwell, M ;
Singh, BM ;
Fortman, C ;
Bauman, BD ;
Lukes, J ;
Gerson, MC .
JOURNAL OF NUCLEAR CARDIOLOGY, 2002, 9 (01) :52-61
[26]   Image reconstruction in regions-of-interest from truncated projections in a reduced fan-beam scan [J].
Zou, Y ;
Pan, XC ;
Sidky, EY .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (01) :13-27