Functional integration of serial dilution and capillary electrophoresis on a PDMS microchip

被引:7
作者
Chang, JK [1 ]
Heo, YS
Bang, H
Cho, K
Chung, S
Chung, C
Han, DC
机构
[1] Seoul Natl Univ, Sch Elect Engn & Comp Sci, Seoul 151742, South Korea
[2] Seoul Natl Univ, Inst Adv Machinery & Design, Digital Bio Technol Co, Seoul 151742, South Korea
[3] Seoul Natl Univ, Sch Mech & Aerosp Engn, Seoul 151742, South Korea
关键词
serial dilution; capillary electrophoresis; microchip; polydimethylsiloxane (PDMS); lab-on-a-chip;
D O I
10.1007/BF02942271
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
For the quantitative analysis of an unknown sample a calibration curve should be obtained, as analytical instruments give relative, rather than absolute measurements. Therefore, researchers should make standard samples with various known concentrations, measure each standard and the unknown sample, and then determine the concentration of the unknown by comparing the measured value to those of the standards. These procedures are tedious and time-consuming. Therefore, we developed a polymer based microfluidic device from polydimethylsiloxane, which integrates serial dilution and capillary electrophoresis functions in a single device. The integrated microchip can provide a one-step analytical tool, and thus replace the complex experimental procedures. Two plastic syringes, one containing a buffer solution and the other a standard solution,,were connected to two inlet holes on a microchip, and pushed by a hydrodynamic force. The standard sample is serially diluted to various concentrations through the microfluidic networks. The diluted samples are sequentially introduced through microchannels by electro-osmotic force, and their laser-induced fluorescence signals measured by capillary electrophoresis. We demonstrate the integrated microchip performance by measuring the fluorescence signals of fluorescein at various concentrations. The calibration curve obtained from the electropherograms showed the expected linearity.
引用
收藏
页码:233 / 239
页数:7
相关论文
共 42 条
[1]  
BANG H, 2002, UNPUB MICROSYSTEM TE
[2]   An integrated nanoliter DNA analysis device [J].
Burns, MA ;
Johnson, BN ;
Brahmasandra, SN ;
Handique, K ;
Webster, JR ;
Krishnan, M ;
Sammarco, TS ;
Man, PM ;
Jones, D ;
Heldsinger, D ;
Mastrangelo, CH ;
Burke, DT .
SCIENCE, 1998, 282 (5388) :484-487
[3]   Development of a multichannel microfluidic analysis system employing affinity capillary electrophoresis for immunoassay [J].
Cheng, SB ;
Skinner, CD ;
Taylor, J ;
Attiya, S ;
Lee, WE ;
Picelli, G ;
Harrison, DJ .
ANALYTICAL CHEMISTRY, 2001, 73 (07) :1472-1479
[4]   Microchip-based capillary electrophoresis for immunoassays: Analysis of monoclonal antibodies and theophylline [J].
Chiem, N ;
Harrison, DJ .
ANALYTICAL CHEMISTRY, 1997, 69 (03) :373-378
[5]   Multiport flow-control system for lab-on-a-chip microfluidic devices [J].
Chien, RL ;
Parce, JW .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 371 (02) :106-111
[6]   Microchip injection and separation anomalies due to pressure effects [J].
Crabtree, HJ ;
Cheong, ECS ;
Tilroe, DA ;
Backhouse, CJ .
ANALYTICAL CHEMISTRY, 2001, 73 (17) :4079-4086
[7]  
Dolník V, 2000, ELECTROPHORESIS, V21, P41, DOI 10.1002/(SICI)1522-2683(20000101)21:1<41::AID-ELPS41>3.0.CO
[8]  
2-7
[9]   INTERNAL STANDARDIZATION TECHNIQUE FOR CAPILLARY ZONE ELECTROPHORESIS [J].
DOSE, EV ;
GUIOCHON, GA .
ANALYTICAL CHEMISTRY, 1991, 63 (11) :1154-1158
[10]   Microfabricated centrifugal microfluidic systems: Characterization and multiple enzymatic assays [J].
Duffy, DC ;
Gillis, HL ;
Lin, J ;
Sheppard, NF ;
Kellogg, GJ .
ANALYTICAL CHEMISTRY, 1999, 71 (20) :4669-4678