Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic-on-ceramic total hip replacement

被引:54
作者
Al-Hajjar, Mazen [1 ]
Leslie, Ian J. [1 ]
Tipper, Joanne [1 ,2 ]
Williams, Sophie [1 ]
Fisher, John [1 ,2 ]
Jennings, Louise M. [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Leeds Teaching Hosp Trust, Leeds Musculoskeletal Biomed Res Unit, Leeds LS2 9JT, W Yorkshire, England
关键词
wear; ceramic on ceramic; hip replacement; microseparation; inclination angle; LONG-TERM WEAR; JOINT SIMULATOR; ACETABULAR CUPS; ALUMINA; METAL; PROSTHESES; BEARING; UHMWPE; ARTHROPLASTY; POLYETHYLENE;
D O I
10.1002/jbm.b.31708
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Ceramic-on-ceramic (CoC) bearings in total hip replacements (THRs) have shown low wear volumes under standard gait in hip simulator studies. However, clinical reports have indicated variations in wear rates and formation of stripe-like wear area on the ceramic femoral heads. The aim of this study was to investigate the influence of cup inclination angle and microseparation on the wear of CoC bearings in THRs. The six station Leeds II Physiological Anatomical Joint Simulator was used to investigate the wear of 28 mm diameter alumina matrix composite ceramic bearings (BIOLOX (R) delta). It was shown that increasing the cup inclination angle from 55 degrees to 65 degrees had no significant effect on the wear rate of BIOLOX (R) delta CoC under both standard gait and microseparation conditions in this in vitro study. Under standard gait conditions, the mean wear rate for both cup inclination angle conditions was very low at 0.05 mm(3)/million cycles. The introduction of microseparation to the standard gait cycle increased-the mean wear rates to 0.13 mm(3)/million cycles for the cup inclination angle of 55 and 0.11 mm(3)/million cycles for that of 65 degrees. The level of increased wear with microseparation was not dependent on cup angle. A stripe of wear on the head also formed, with corresponding superior rim wear on the cup. The wear rates obtained were low compared to the HIPed third generation alumina ceramic (BIOLOX (R) forte) tested under the same adverse conditions (1.84 mm(3)/million cycles). BIOLOX (R) delta has shown lower wear than previous ceramic materials used in THR under adverse conditions. (C) 2010 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 9513: 263-268, 2010.
引用
收藏
页码:263 / 268
页数:6
相关论文
共 28 条
[1]
The influence of implant position on the wear of alumina-on-alumina studied in a hip simulator [J].
Affatato, S ;
Bersaglia, G ;
Foltran, I ;
Emiliani, D ;
Traina, F ;
Toni, A .
WEAR, 2004, 256 (3-4) :400-405
[2]
Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant [J].
Bigsby, RJA ;
Hardaker, CS ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (03) :265-269
[3]
Effect of ion implantation on the tribology of metal-on-metal hip prostheses [J].
Bowsher, JG ;
Hussain, A ;
Williams, P ;
Nevelos, J ;
Shelton, JC .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :107-111
[4]
CLARKE IC, 2006, SEVERE SIMULATION TE, P11
[5]
In vivo determination of hip joint separation and the forces generated due to impact loading conditions [J].
Dennis, DA ;
Komistek, RD ;
Northcut, EJ ;
Ochoa, JA ;
Ritchie, A .
JOURNAL OF BIOMECHANICS, 2001, 34 (05) :623-629
[6]
Hip simulator wear comparison of metal-on-metal, ceramic-on-ceramic and crosslinked UHMWPE bearings [J].
Essner, A ;
Sutton, K ;
Wang, A .
WEAR, 2005, 259 :992-995
[7]
Presidential guest lecture - Tribology of alternative beatings [J].
Fisher, John ;
Jin, Zhongmin ;
Tipper, Joanne ;
Stone, Martin ;
Ingham, Eileen .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (453) :25-34
[8]
Wear and biological activity of highly crosslinked polyethylene in the hip under low serum protein concentrations [J].
Galvin, A. L. ;
Tipper, I. L. ;
Jennings, L. M. ;
Stone, M. H. ;
Jin, Z. M. ;
Ingham, E. ;
Fisher, J. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H1) :1-10
[9]
Clicking and Squeaking: In Vivo Correlation of Sound and Separation for Different Bearing Surfaces [J].
Glaser, Diana ;
Komistek, Richard D. ;
Cates, Harold E. ;
Mahfouz, Mohamed R. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A :112-120
[10]
Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup [J].
Hasegawa, M ;
Sudo, A ;
Hirata, H ;
Uchida, A .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :658-661