Performance of advanced a-Si/CsI-based flat panel x-ray detectors for mammography

被引:23
作者
Albagli, D [1 ]
Hudspeth, H [1 ]
Possin, G [1 ]
Lee, J [1 ]
Granfors, P [1 ]
Giambattista, B [1 ]
机构
[1] GE Global Res Ctr, Niskayuna, NY 12309 USA
来源
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2003年 / 5030卷
关键词
amorphous silicon; flat-panel detector; detective quantum efficiency; DQE; full field digital mammography; FFDM;
D O I
10.1117/12.480010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The GE Senographe 2000D(TM), the first full field digital mammography system based on amorphous Silicon (a-Si) flat panel arrays and a Cesium-Iodide (CsI) scintillator, has been in clinical use for several years. The purpose of this paper is to demonstrate and quantify improvements in the detective quantum efficiency (DQE) for both typical screening and ultra-low exposure levels for this technology platform. A new figure of merit, the electronic noise factor, is introduced to explicitly quantify the influence of the electronic noise, conversion factor, modulation transfer function (MTF), and pixel pitch towards the reduction of DQE at low exposure levels. Methods to improve the DQE through an optimization of both the flat panel design and the scintillator deposition process ate discussed. The results show a substantial improvement in the DQE(f) at all frequencies and demonstrate the potential for DQE(0) to exceed 80%. The combination of high DQE at. ultra low exposures and the inherent fast read-out capability makes this technology platform ideal for both current clinical procedures and advanced applications that may use multiple projections (tomosynthesis) or contrast media to enhance digital mammography.
引用
收藏
页码:553 / 563
页数:11
相关论文
共 9 条
[1]   Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications [J].
Antonuk, LE ;
Jee, KW ;
El-Mohri, Y ;
Maolinbay, M ;
Nassif, S ;
Rong, X ;
Zhao, Q ;
Siewerdsen, JH ;
Street, RA ;
Shah, KS .
MEDICAL PHYSICS, 2000, 27 (02) :289-306
[2]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[3]   Performance of a 41X41-cm2 amorphous silicon flat panel x-ray detector for radiographic imaging applications [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL PHYSICS, 2000, 27 (06) :1324-1331
[4]   DQE(f) of an amorphous silicon flat panel x-ray detector: Detector parameter influences and measurement methodology [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :2-13
[5]  
Hubbell J.H., 1997, Tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients (version 1.03), DOI DOI 10.18434/T4D01F
[6]   The x-ray sensitivity of amorphous selenium for mammography [J].
Stone, MF ;
Zhao, W ;
Jacak, BV ;
O'Connor, P ;
Yu, B ;
Rehak, P .
MEDICAL PHYSICS, 2002, 29 (03) :319-324
[7]   Modeling the x-ray energy characteristics of DQE for full field digital mammography [J].
Tkaczyk, JE ;
LeBlanc, J ;
Nevin, R ;
Kautz, G ;
Albagli, D ;
Sandrik, J ;
Granfors, P .
MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 :570-581
[8]   Full breast digital mammography with an amorphous silicon-based flat panel detector: Physical characteristics of a clinical prototype [J].
Vedantham, S ;
Karellas, A ;
Suryanarayanan, S ;
Albagli, D ;
Han, S ;
Tkaczyk, EJ ;
Landberg, CE ;
Opsahl-Ong, B ;
Granfors, PR ;
Levis, I ;
D'Orsi, CJ ;
Hendrick, RE .
MEDICAL PHYSICS, 2000, 27 (03) :558-567
[9]   Characterization of a full field digital mammography detector based on direct x-ray conversion in selenium [J].
Yorker, JG ;
Jeromin, LS ;
Lee, DLY ;
Palecki, EF ;
Golden, KP ;
Jing, ZX .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :21-29