Effects of sterilization on the Tekscan digital pressure sensor

被引:27
作者
Agins, HJ
Harder, VS
Lautenschlager, EP
Kudrna, JC
机构
[1] Illinois Bone & Joint Inst, Glenview, IL 60025 USA
[2] Evanston Norhtwestern Healthcare, Dept Orthopaed Surg, Evanston, IL 60201 USA
[3] Northwestern Univ, Orthopaed Biomat Lab, Chicago, IL 60611 USA
关键词
sterilization; Tekscan; orthopaedic biomechanics; calibration;
D O I
10.1016/S1350-4533(03)00119-X
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Investigations into the effects of sterilization on a new biomechanical pressure sensor are necessary before contemplating in vivo use. Ten. designated Experimental, "K-Scan" digital pressure sensor arrays were sterilized with ethylene oxide gas (EtO), and their ability to accurately and reproducibly measure an applied load of 2225 N (500 lb) was assessed. Simultaneously, 10 un-sterilized sensor arrays. designated Control, were assessed. Each array was loaded 10 times inside a two-dimensional curved surface, and all arrays exhibited high reproducibility (coefficients of variation < 2.0%). Following sterilization, the Experimental sensors showed a 22.2% average decrease in recorded force, a statistically significant difference from the pre-sterile data (p < 0.002). However, when the Experimental sensors were re-calibrated post-sterilization, they showed only a 0.1% average decrease in recorded force, not a statistically significant difference (p > 0.05, beta < 0.05). Following 1-week storage, trial 2 data of the Control sensors showed a less dramatic yet significant 3.4% average decrease in recorded force when compared to trial 1 data (p < 0.02). Control trial 2, once re-calibrated. showed a 0.5% average decrease in recorded force, not a statistically significant difference (p > 0.05, beta < 0.05). Results suggest that, following EtO sterilization, accurate and reproducible pressure measurements can be obtained from K-Scan sensors when calibration is performed at time of use. (C) 2003 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:775 / 780
页数:6
相关论文
共 31 条
[1]
Contact mechanics of the medial tibial plateau after implantation of a medial meniscal allograft - A human cadaveric study [J].
Alhalki, MM ;
Hull, ML ;
Howell, SM .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (03) :370-376
[2]
MENISCAL TEARS - THE EFFECT OF MENISCECTOMY AND OF REPAIR ON INTRAARTICULAR CONTACT AREAS AND STRESS IN THE HUMAN KNEE - A PRELIMINARY-REPORT [J].
BARATZ, ME ;
FU, FH ;
MENGATO, R .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1986, 14 (04) :270-275
[3]
Statically equivalent load and support conditions produce different hip joint contact pressures and periacetabular strains [J].
Bay, BK ;
Hamel, AJ ;
Olson, SA ;
Sharkey, NA .
JOURNAL OF BIOMECHANICS, 1997, 30 (02) :193-196
[4]
PRESSURE DISTRIBUTION AT THE KNEE-JOINT - INFLUENCE OF VARUS AND VALGUS DEVIATION WITHOUT AND WITH LIGAMENT DISSECTION [J].
BRUNS, J ;
VOLKMER, M ;
LUESSENHOP, S .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1993, 113 (01) :12-19
[5]
PRESSURE DISTRIBUTION IN THE KNEE-JOINT - INFLUENCE OF FLEXION WITH AND WITHOUT LIGAMENT DISSECTION [J].
BRUNS, J ;
VOLKMER, M ;
LUESSENHOP, S .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1994, 113 (04) :204-209
[6]
Calcaneocuboid joint pressures with lateral column lengthening (Evans) procedure [J].
Cooper, PS ;
Nowak, MD ;
Shaer, J .
FOOT & ANKLE INTERNATIONAL, 1997, 18 (04) :199-205
[7]
THE CONTACT AREA AND PRESSURE DISTRIBUTION PATTERN OF THE KNEE - A STUDY OF NORMAL AND OSTEOARTHROTIC KNEE JOINTS [J].
FUKUBAYASHI, T ;
KUROSAWA, H .
ACTA ORTHOPAEDICA SCANDINAVICA, 1980, 51 (06) :871-879
[8]
CONTACT STRESS GRADIENT DETECTION LIMITS OF PRESSENSOR FILM [J].
HALE, JE ;
BROWN, TD .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1992, 114 (03) :352-357
[9]
An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji film [J].
Harris, ML ;
Morberg, P ;
Bruce, WJM ;
Walsh, WR .
JOURNAL OF BIOMECHANICS, 1999, 32 (09) :951-958
[10]
THE NORMAL GLENOHUMERAL RELATIONSHIPS - AN ANATOMICAL STUDY OF 140 SHOULDERS [J].
IANNOTTI, JP ;
GABRIEL, JP ;
SCHNECK, SL ;
EVANS, BG ;
MISRA, S .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (04) :491-500