Development and calibration of a load sensing cervical distractor capable of withstanding autoclave sterilization

被引:8
作者
Demetropoulos, CK
Truumees, E
Herkowitz, HN
Yang, KH
机构
[1] Wayne State Univ, Ctr Bioengn, Detroit, MI 48201 USA
[2] William Beaumont Hosp, Res Inst, Ctr Biomech & Implant Anal, Royal Oak, MI 48073 USA
关键词
cervical distractor; spinal fusion; cervical discectomy; strain gauge;
D O I
10.1016/j.medengphy.2004.09.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
In surgery of the cervical spine, a Caspar pin distractor is often used to apply a tensile load to the spine in order to open up the disc space. This is often done in order to place a graft or other interbody fusion device in the spine. Ideally a tight interference fit is achieved. If the spine is over distracted, allowing for a large graft, there is an increased risk of subsidence into the endplate. If there is too little distraction, there is an increased risk of graft dislodgement or pseudoarthrosis. Generally, graft height is selected from preoperative measurements and observed distraction without knowing the intraoperative compressive load. This device was designed to give the surgeon an assessment of this applied load. Instrumentation of the device involved the application of strain gauges and the selection of materials that would survive standard autoclave sterilization. The device was calibrated, sterilized and once again calibrated to demonstrate its suitability for surgical use. Results demonstrate excellent linearity in the calibration, and no difference was detected in the pre- and post-sterilization calibrations. (c) 2004 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:343 / 346
页数:4
相关论文
共 8 条
[1]
BROWER RS, 1992, EFFECT DISTRACTION U
[2]
A proposed technique for intraoperative measurement of cervical spine stiffness [J].
Frank, E ;
Chamberland, D ;
Ragel, B .
NEUROSURGERY, 1996, 39 (01) :147-150
[3]
BIOMECHANICAL ANALYSIS OF FACET AND GRAFT LOADING IN A SMITH-ROBINSON TYPE CERVICAL-SPINE MODEL [J].
OLSEWSKI, JM ;
GARVEY, TA ;
SCHENDEL, MJ .
SPINE, 1994, 19 (22) :2540-2544
[4]
The Technical Staff of Measurements Group Inc., 1988, STRAIN GAG BAS TRANS
[5]
Effects of a cervical compression plate on graft forces in an anterior cervical discectomy model [J].
Truumees, E ;
Demetropoulos, CK ;
Yang, KH ;
Herkowitz, HN .
SPINE, 2003, 28 (11) :1097-1102
[6]
Effects of disc height and distractive forces on graft compression in an anterior cervical discectomy model [J].
Truumees, E ;
Demetropoulos, CK ;
Yang, KH ;
Herkowitz, HN .
SPINE, 2002, 27 (22) :2441-2445
[7]
TRUUMEES E, 2000, INSTRUCTIONAL COURSE, V49
[8]
Truumees E, 1999, SEMIN SPINE SURG, V11, P373