Cyclic load testing and ultimate failure strength of biodegradable glenoid anchors

被引:47
作者
Barber, F. Alan [1 ]
Coons, David A. [2 ]
Ruiz-Suarez, Michell [3 ]
机构
[1] Plano Orthoped & Sports Med Ctr, Plano, TX 75093 USA
[2] Pacific Sport Med, Tacoma, WA USA
[3] Natl Inst Rehabil, Mexico City, DF, Mexico
关键词
arthroscopy; bankart; biodegradable; biomechanics; cyclic loading; suture anchor;
D O I
10.1016/j.arthro.2007.08.011
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: The purpose of this study was to compare biodegradable glenoid suture anchors by cyclic loading and load to failure testing. Methods: Seven different suture anchors (BioKnotless and Lupine Loop [DePuy-Mitek, Norwood, MA]; BioPushLock, BioSutureTak, and BioFasTak [Arthrex Corp, Naples, FL]; BioAnchor [Conmed Linvatec, Largo FL]; and BioRaptor [Smith & Nephew, Andover, MA]) were tested in 8 matched pairs of human cadaver fresh-frozen glenoids. The anchors were inserted in rotation into different glenoid rim positions. Sutures attached to the anchors were fixed to an Instron 8871 machine (Instron, Canton, MA) and cyclic loading and destructive testing were performed. The cyclic displacement at 100 and 500 cycles, stiffness, ultimate failure strength, and mode of failure were determined. Results: No statistical difference was found in the ultimate failure load for any of these anchors. The Lupine Loop and BioAnchor had greater 100 cycle and 500 cycle mean displacements than the BioPushLock and BioSutureTak. The Lupine Loop also had greater 100 cycle and 500 cycle mean displacement than the BioFasTak (P < .05). The BioAnchor had greater mean 500 cyclic displacement than the BioFasTak (P < .05). Mean BioSutureTak stiffness was greater than the Lupine Loop, BioAnchor, BioKnotless, and BioRaptor (P < .05). Conclusions: No differences in ultimate failure strength after cyclic loading were found in these seven biodegradable glenoid anchors (BioKnotless, Lupine Loop, BioPushLock, BioSutureTak, BioFasTak, BioAnchor, and BioRaptor). Most displacement occurred in the first 100 cycles. Displacement at 500 cycles was greater for the Lupine Loop and the BioAnchor than the BioPushLock, BioSutureTak, and BioFasTak. Failure was principally by the anchor pulling out of bone except for the BioSutureTak, which also failed by the suture loop eyelet pulling out of the anchor body, and the BioPushLock which failed by the suture slipping past anchor. Clinical Relevance: Biodegradable glenoid anchors did not show statistical difference in ultimate failure load after cyclic loading.
引用
收藏
页码:224 / 228
页数:5
相关论文
共 25 条
[1]
Cyclic load testing of biodegradable suture anchors containing 2 high-strength sutures [J].
Barber, F. Alan ;
Coons, David A. ;
Ruiz-Suarez, Michell .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2007, 23 (04) :355-360
[2]
Biodegradable shoulder anchors have unique modes of failure [J].
Barber, F. Alan .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2007, 23 (03) :316-320
[3]
Sutures and suture anchors - Update 2006 [J].
Barber, F. Alan ;
Herbert, Morley A. ;
Coons, David A. ;
Boothby, Michael H. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (10) :1063-1069
[4]
The relationship of suture anchor failure and bone density to proximal humerus location: A cadaveric study [J].
Barber, FA ;
Feder, SM ;
Burkhart, SS ;
Ahrens, J .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 1997, 13 (03) :340-345
[5]
Cyclic loading of transosseous rotator cuff repairs: Tension overload as a possible cause of failure [J].
Burkhart, SS ;
Johnson, TC ;
Wirth, MA ;
Athanasiou, KA .
ARTHROSCOPY, 1997, 13 (02) :172-176
[6]
Clinical results of arthroscopic Bankart repair with knot-tying and knotless suture anchors [J].
Cho, Nam Su ;
Lubis, Andri M. T. ;
Ha, Jeong Han ;
Rhee, Yong Girl .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (12) :1276-1282
[7]
Mode of failure for rotator cuff repair with suture anchors identified at revision surgery [J].
Cummins, CA ;
Murrell, GAC .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2003, 12 (02) :128-133
[8]
Rotator cuff repair: An ex vivo analysis of suture anchor repair techniques on initial load to failure [J].
Cummins, CA ;
Appleyard, RC ;
Strickland, S ;
Haen, PS ;
Chen, SY ;
Murrell, GAC .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2005, 21 (10) :1236-1241
[9]
Biomechanical comparison of 3 suture anchor configurations for repair of type II SLAP lesions [J].
Domb, Benjamin G. ;
Ehteshami, John R. ;
Shindle, Michael K. ;
Gulotta, Lawrence ;
Zoghi-Moghadam, Mohamad ;
MacGillivray, John D. ;
Altchek, David W. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2007, 23 (02) :135-140
[10]
All arthroscopic rotator cuff repairs (Reprinted from Operative Techniques in Sports Medicine, October, 2000) [J].
Gartsman, GM .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2001, 32 (03) :501-+