On-line monitoring of glucose and penicillin by sequential injection analysis

被引:34
作者
Min, RW
Nielsen, J
Villadsen, J
机构
[1] TECH UNIV DENMARK,CTR PROC BIOTECHNOL,DK-2800 LYNGBY,DENMARK
[2] TECH UNIV DENMARK,DEPT BIOTECHNOL,DK-2800 LYNGBY,DENMARK
关键词
enzymatic methods; chemiluminescence; glucose; penicillin; on-line monitoring; cultivation; sequential injection analysis;
D O I
10.1016/0003-2670(95)00523-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sequential injection analysis (SIA) system has been developed for on-line monitoring of glucose and penicillin during cultivations of the filamentous fungus Penicillium chrysogenum. The SIA system consists of a peristaltic pump, an injection valve, two piston pumps, two multi-position valves and a detector. The glucose analyzer is based on an enzymatic reaction using glucose oxidase, which converts glucose to glucono-lactone with formation of hydrogen peroxide and subsequent detection of H2O2 by a chemiluminescence reaction involving luminol. The penicillin analysis is based on formation of penicilloic acid by penicillinase. Penicilloic acid is detected either by a chemiluminescence method or by a decolorization method. In the chemiluminescence method the penicilloic acid is quantified by its quenching effect on the chemiluminescence signal obtained when luminol reacts with iodine. In the decolorization method the penicilloic acid is detected spectrophotometrically by the decrease in the absorbance of an iodine-starch complex.
引用
收藏
页码:199 / 205
页数:7
相关论文
共 41 条
[21]   RAPID MEASUREMENT OF PENICILLIN CONTAINED IN COMPLEX MEDIA USING ENZYME-LOADED GLASS ELECTRODES [J].
MEIER, H ;
KUMARAN, S ;
DANNA, AM ;
TRANMINH, C .
ANALYTICA CHIMICA ACTA, 1991, 249 (02) :405-411
[22]   APPLICATION AND AUTOMATION OF FLOW-INJECTION ANALYSIS (FIA) USING FAST RESPONDING ENZYME GLASS ELECTRODES TO DETECT PENICILLIN IN FERMENTATION BROTH AND UREA IN HUMAN SERUM [J].
MEIER, H ;
LANTREIBECQ, F ;
TRANMINH, C .
JOURNAL OF AUTOMATIC CHEMISTRY, 1992, 14 (04) :137-143
[23]   CULTURE FLUORESCENCE MEASUREMENTS DURING BATCH AND FED-BATCH CULTIVATIONS WITH PENICILLIUM-CHRYSOGENUM [J].
NIELSEN, J ;
JOHANSEN, CL ;
VILLADSEN, J .
JOURNAL OF BIOTECHNOLOGY, 1994, 38 (01) :51-62
[24]  
Nielsen J., 1992, Process Control and Quality, V2, P371
[25]   FIA FOR ONLINE MONITORING OF IMPORTANT LACTIC-ACID FERMENTATION VARIABLES [J].
NIELSEN, J ;
NIKOLAJSEN, K ;
VILLADSEN, J .
BIOTECHNOLOGY AND BIOENGINEERING, 1989, 33 (09) :1127-1134
[26]   IN-LINE FLOW-INJECTION ANALYSIS FOR MONITORING LACTIC-ACID FERMENTATIONS [J].
NIKOLAJSEN, K ;
NIELSEN, J ;
VILLADSEN, J .
ANALYTICA CHIMICA ACTA, 1988, 214 (1-2) :137-145
[27]   SIMPLE AND MILD PREPARATION OF AN ENZYME-IMMOBILIZED MEMBRANE FOR A BIOSENSOR USING BETA-TYPE CRYSTALLINE CHITIN [J].
OHASHI, E ;
KORIYAMA, T .
ANALYTICA CHIMICA ACTA, 1992, 262 (01) :19-25
[28]   A FLOW-INJECTION SYSTEM BASED ON IMMOBILIZED PENICILLINASE AND A LINEAR PH-BUFFER FOR POTENTIOMETRIC DETERMINATION OF PENICILLIN [J].
OLSSON, B .
ANALYTICA CHIMICA ACTA, 1988, 209 (1-2) :123-133
[29]  
PETERSON JI, 1979, ANAL CHIM ACTA, V16, P207
[30]   INDUSTRIAL ONLINE MONITORING OF PENICILLIN-V, GLUCOSE AND ETHANOL USING A SPLIT-FLOW MODIFIED THERMAL BIOSENSOR [J].
RANK, M ;
GRAM, J ;
DANIELSSON, B .
ANALYTICA CHIMICA ACTA, 1993, 281 (03) :521-526