Assessment of a customised immobilisation system for head and neck IMRT using electronic portal imaging

被引:59
作者
Humphreys, M
Urbano, MTG
Mubata, C
Mites, E
Harrington, KJ
Bidmead, M
Nutting, CM
机构
[1] Royal Marsden NHS Trust, Dept Radiotherapy, London, England
[2] Royal Marsden NHS Trust, Inst Canc Res, Head & Neck Unit, London, England
[3] Royal Marsden NHS Trust, Dept Phys, London, England
关键词
immobilisation; head and neck; IMRT; electronic portal imaging;
D O I
10.1016/j.radonc.2005.06.039
中图分类号
R73 [肿瘤学];
学科分类号
100214 [肿瘤学];
摘要
Purpose: To evaluate set-up reproducibility of a cabulite shell and determine CTV-PTV margins for head and neck intensity-modulated-radiotherapy. Materials and methods: Twenty patients were entered into the study. A total of 354 anterior and lateral isocentric electronic portal images (EPIs) were compared to simulator reference images. Results: About 94% of all translational displacements were <= 3 mm, and 99% <= 5 mm. The overall systematic error was 0.9 mm (+/- 1.0SD) in the Right-Left, 0.7 mm (+/- 0.90) in the Superior-inferior and -0.02 mm (+/- 1.1SD) in the Anterior-Posterior directions. The corresponding SDs of the random errors were +/- 0.4, +/- 0.6 and +/- 0.7 mm. The estimated margins required from CTV-PTV were calculated according to the Van Herk formula was 2.9, 2.6 and 3.3 mm, respectively. Conclusions: This head and neck immobilisation system is of sufficient accuracy for its use with IMRT treatments and a 3 mm CTV-PTV margin has been adopted. (C) 2005 Published by Elsevier Ireland Ltd.
引用
收藏
页码:39 / 44
页数:6
相关论文
共 18 条
[1]
SETUP DEVIATIONS IN WEDGED PAIR IRRADIATION OF PAROTID-GLAND AND TONSILLAR TUMORS, MEASURED WITH AN ELECTRONIC PORTAL IMAGING DEVICE [J].
BEL, A ;
KEUS, R ;
VIJLBRIEF, RE ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (02) :153-159
[2]
Comparison of two dead and neck immobilization systems [J].
Bentel, GC ;
Marks, LB ;
Hendren, K ;
Brizel, DM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :867-873
[3]
Bidmead M, 2003, GEOMETRIC UNCERTAINT
[4]
Electronic portal image assisted reduction of systematic set-up errors in head and neck irradiation [J].
de Boer, HCJ ;
de Koste, JRV ;
Creutzberg, CL ;
Visser, AG ;
Levendag, PC ;
Heijmen, BJM .
RADIOTHERAPY AND ONCOLOGY, 2001, 61 (03) :299-308
[5]
Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors [J].
Gilbeau, L ;
Octave-Prignot, M ;
Loncol, T ;
Renard, L ;
Scalliet, P ;
Grégoire, V .
RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) :155-162
[6]
ACCURACY OF FIELD ALIGNMENT IN RADIOTHERAPY OF HEAD AND NECK-CANCER UTILIZING INDIVIDUALIZED FACE MASK IMMOBILIZATION - A RETROSPECTIVE ANALYSIS OF CLINICAL-PRACTICE [J].
HESS, CF ;
KORTMANN, RD ;
JANY, R ;
HAMBERGER, A ;
BAMBERG, M .
RADIOTHERAPY AND ONCOLOGY, 1995, 34 (01) :69-72
[7]
ACCURACY IN RADIATION-FIELD ALIGNMENT IN HEAD AND NECK-CANCER - A PROSPECTIVE-STUDY [J].
HUIZENGA, H ;
LEVENDAG, PC ;
DEPORRE, PMZR ;
VISSER, AG .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (02) :181-187
[8]
THE EFFECT OF SETUP UNCERTAINTIES ON THE TREATMENT OF NASOPHARYNX CANCER [J].
HUNT, MA ;
KUTCHER, GJ ;
BURMAN, C ;
FASS, D ;
HARRISON, L ;
LEIBEL, S ;
FUKS, Z .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (02) :437-447
[9]
Set-up verification using portal imaging; review of current clinical practice [J].
Hurkmans, CW ;
Remeijer, P ;
Lebesque, JV ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) :105-120
[10]
*INT COMM RAD UN M, ICRU62