Motion correction algorithms may create spurious brain activations in the absence of subject motion

被引:245
作者
Freire, L [1 ]
Mangin, JF
机构
[1] Serv Hosp Frederic Joliot, CEA, F-91401 Orsay, France
[2] Inst Med Nucl, FML, P-1649028 Lisbon, Portugal
关键词
fMRI; motion correction; artifact; spurious activation; robust registration;
D O I
10.1006/nimg.2001.0869
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
This paper describes several experiments that prove that standard motion correction methods may induce spurious activations in some motion-free fMRI studies. This artifact stems from the fact that activated areas behave like biasing outliers for the difference of square-based measures usually driving such registration methods. This effect is demonstrated first using a motion-free simulated time series including artificial activation-like signal changes. Several additional simulations explore the influence of activation amplitude and extent. The effect is finally highlighted on an actual time series obtained from a 3-T magnet. All the experiments are performed using four different realignment methods, which allows us to show that the problem may be overcome by methods based on a robust similarity measure like mutual information. (C) 2001 Academic Press.
引用
收藏
页码:709 / 722
页数:14
相关论文
共 33 条
[1]   How to estimate global activity independent of changes in local activity [J].
Andersson, JLR .
NEUROIMAGE, 1997, 6 (04) :237-244
[2]  
BIRN RM, 1997, P ISMRM 5 ANN M VANC, P1913
[3]   Movement-related effects in fMRI time-series [J].
Friston, KJ ;
Williams, S ;
Howard, R ;
Frackowiak, RSJ ;
Turner, R .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) :346-355
[4]   Spatial registration and normalization of images [J].
Friston, KJ ;
Ashburner, J ;
Frith, CD ;
Poline, JB ;
Heather, JD ;
Frackowiak, RSJ .
HUMAN BRAIN MAPPING, 1995, 3 (03) :165-189
[5]  
FROUIN V, 1997, HUM BRAIN MAPP, V5, pS458
[6]   Characterization and correction of interpolation effects in the realignment of fMRI time series [J].
Grootoonk, S ;
Hutton, C ;
Ashburner, J ;
Howseman, AM ;
Josephs, O ;
Rees, G ;
Friston, KJ ;
Turner, R .
NEUROIMAGE, 2000, 11 (01) :49-57
[7]   ARTIFACTS DUE TO STIMULUS CORRELATED MOTION IN FUNCTIONAL IMAGING OF THE BRAIN [J].
HAJNAL, JV ;
MYERS, R ;
OATRIDGE, A ;
SCHWIESO, JE ;
YOUNG, IR ;
BYDDER, GM .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (03) :283-291
[8]   Medical image registration [J].
Hill, DLG ;
Batchelor, PG ;
Holden, M ;
Hawkes, DJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03) :R1-R45
[9]   Voxel similarity measures for 3-D serial MR brain image registration [J].
Holden, M ;
Hill, DLG ;
Denton, ERE ;
Jarosz, JM ;
Cox, TCS ;
Rohlfing, T ;
Goodey, J ;
Hawkes, DJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (02) :94-102
[10]  
Jezzard P, 1999, HUM BRAIN MAPP, V8, P80, DOI 10.1002/(SICI)1097-0193(1999)8:2/3<80::AID-HBM2>3.0.CO