Development of standard operation procedures for the manufacture of n-octadecyl bonded silicas as packing material in certified reference columns for reversed-phase liquid chromatography

被引:30
作者
von Hohenesche, CD
Ehwald, V
Unger, KK
机构
[1] Johannes Gutenberg Univ Mainz, Inst Inorgan Chem & Analyt Chem, D-55099 Mainz, Germany
[2] Wacker Chem GmbH, D-87416 Kempten, Germany
关键词
standard operation procedure; stationary phases; LC; reference material; silica; n-octadecyl bonded;
D O I
10.1016/j.chroma.2003.10.091
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The development of standard operation procedures for the manufacture of a n-octadecyl bonded spherical silica packing from partially condensed tetraethoxysilane as silica source is described. The synthesis comprises five intermediate products and six synthesis steps which were examined according to their reproducibility and robustness. The results led to the optimisation of the manufacturing process for a n-octadecyl bonded silica. Correlations were drawn between the dynamic viscosity of the poly(ethoxy)siloxane (PES), the synthesis parameters, the resulting pore structural properties and particle size distribution of the silicas. Validated procedures were developed to manufacture spherical porous ultra-pure silicas with a specific surface area of 350 m(2) g(-1) +/-5% R.S.D., a specific pore volume of 1.0 ml g(-1) +/-3.7% R.S.D., an average pore diameter of 12.0 nm +/-0.5% R.S.D. and an average particle diameter of 5 mum. Results are presented on trial batches and the final master batch which were both used as packing materials in reversed-phase liquid chromatography (RP-LC) columns. The latter columns were certified and accepted as an HPLC column as reference material (BCR-722) by the European Commission, Institute for Reference Materials and Measurements (IRMM), Geel, Belgium. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:177 / 187
页数:11
相关论文
共 24 条
[1]  
EHWALD V, 1998, THESIS J GUTENBERG U
[2]   Impact of the post-treatment conditions of parent silica on the silanization of n-octadecyl bonded silica packings in reversed-phase high-performance liquid chromatography [J].
Grossmann, F ;
Ehwald, V ;
von Hohenesche, CD ;
Unger, KK .
JOURNAL OF CHROMATOGRAPHY A, 2001, 910 (02) :223-236
[3]   On the need to retain unregulated choice of high-performance liquid chromatography columns [J].
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 1999, 852 (02) :603-605
[4]  
HEARN MTW, 1997, EU STANDARD HPLC COL, pA8
[5]   DETERMINATION OF HYDROXYL-GROUPS AND WATER-CONTENT IN SILICA BY NUCLEAR MAGNETIC-RESONANCE SPECTROSCOPY [J].
HOLIK, M ;
MATEJKOVA, B .
JOURNAL OF CHROMATOGRAPHY, 1981, 213 (01) :33-39
[6]  
Iler RK, 1955, The colloid chemistry of silica and silicatesvol, V80
[7]   Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns III.: Results obtained with Kromasil C18 columns [J].
Kele, M ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 1999, 855 (02) :423-453
[8]   Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns -: IV.: Results obtained with luna C18 (2) columns [J].
Kele, M ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2000, 869 (1-2) :181-209
[9]   Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatographic columns II.: Results obtained with Symmetry C18 columns [J].
Kele, M ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 1999, 830 (01) :55-79
[10]   IMPROVED SILICA-BASED COLUMN PACKINGS FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
KOHLER, J ;
KIRKLAND, JJ .
JOURNAL OF CHROMATOGRAPHY, 1987, 385 :125-150