A 16-element phased-array head coil

被引:62
作者
Porter, JR [1 ]
Wright, SM [1 ]
Reykowski, A [1 ]
机构
[1] Texas A&M Univ, Dept Elect Engn, College Stn, TX 77843 USA
关键词
phased-array coils; surface coils; MRI; time domain multiplexing;
D O I
10.1002/mrm.1910400213
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Volume-array coils offer increased signal-to-noise ratio (SNR) over standard volume coils near the array elements while preserving the SNR at the center of the volume. As the number of array elements is increased, the SNR advantage as well as the complexity of actually constructing the array increases also. In this study, a 16-channel receive-only array for imaging of the brain is demonstrated and compared to a circularly polarized (CP) head coil of similar shape and diameter. The array was formed from a 2 x 8 grid of square elements placed on a cylindrical form. Mutual coupling was minimized by a combination of overlapping element placement and current-reducing matching networks. Simultaneous data acquisition from the 16 individual elements was performed using a four-channel receiver system with each channel time domain multiplexed by a factor of 4, Theoretical and experimental comparisons between the array and a standard CP head coil show that the array offers an increase in SNR of nearly a factor of 3 near its surface while maintaining a comparable SNR to that of the CP head coil in the center of the region of interest.
引用
收藏
页码:272 / 279
页数:8
相关论文
共 24 条
[1]   IMPROVED SURFACE COIL IMAGING IN MR - DECOUPLING OF THE EXCITATION AND RECEIVER COILS [J].
BOSKAMP, EB .
RADIOLOGY, 1985, 157 (02) :449-452
[2]   What is the optimum phased array coil design for cardiac and torso magnetic resonance? [J].
Bottomley, PA ;
Olivieri, CHL ;
Giaquinto, R .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) :591-599
[3]   AN NMR PHASED-ARRAY FOR HUMAN CARDIAC P-31 SPECTROSCOPY [J].
HARDY, CJ ;
BOTTOMLEY, PA ;
ROHLING, KW ;
ROEMER, PB .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) :54-64
[4]   VOLUME IMAGING WITH MR PHASED-ARRAYS [J].
HAYES, CE ;
HATTES, N ;
ROEMER, PB .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (02) :309-319
[5]   NOISE CORRELATIONS IN DATA SIMULTANEOUSLY ACQUIRED FROM MULTIPLE SURFACE COIL ARRAYS [J].
HAYES, CE ;
ROEMER, PB .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (02) :181-191
[6]   PELVIC IMAGING WITH PHASED-ARRAY COILS - QUANTITATIVE ASSESSMENT OF SIGNAL-TO-NOISE RATIO IMPROVEMENT [J].
HAYES, CE ;
DIETZ, MJ ;
KING, BF ;
EHMAN, RL .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (03) :321-326
[7]   FAST MRI DATA ACQUISITION USING MULTIPLE DETECTORS [J].
HUTCHINSON, M ;
RAFF, U .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (01) :87-91
[8]   A DECOUPLED COIL DETECTOR ARRAY FOR FAST IMAGE ACQUISITION IN MAGNETIC-RESONANCE-IMAGING [J].
KWIAT, D ;
EINAV, S ;
NAVON, G .
MEDICAL PHYSICS, 1991, 18 (02) :251-265
[9]   PHASED-ARRAY COILS FOR UPPER EXTREMITY MRA [J].
MONROE, JW ;
SCHMALBROCK, P ;
SPIGOS, DG .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (02) :224-229
[10]   A 4-CHANNEL TIME-DOMAIN MULTIPLEXER - A COST-EFFECTIVE ALTERNATIVE TO MULTIPLE RECEIVERS [J].
PORTER, JR ;
WRIGHT, SM ;
FAMILI, N .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) :499-504