A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac Output

被引:407
作者
Critchley, Lester A. [1 ]
Lee, Anna [1 ]
Ho, Anthony M. -H. [1 ]
机构
[1] Chinese Univ Hong Kong, Prince Wales Hosp, Dept Anaesthesia & Intens Care, Shatin, Hong Kong, Peoples R China
关键词
ARTERIAL-PRESSURE WAVE; THORACIC ELECTRICAL BIOIMPEDANCE; BOLUS THERMODILUTION TECHNIQUE; PULSE CONTOUR; FORM ANALYSIS; IMPEDANCE CARDIOGRAPHY; ESOPHAGEAL DOPPLER; TRANSESOPHAGEAL ECHOCARDIOGRAPHY; TRANSPULMONARY THERMODILUTION; LUNG FLUID;
D O I
10.1213/ANE.0b013e3181f08a5b
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Numerous cardiac output (CO) monitors have been produced that provide continuous rather than intermittent readings. Bland and Altman has become the standard method for validating their performance against older standards. However, the Bland and Altman method only assesses precision and does not assess how well a device detects serial changes in CO (trending ability). Currently, there is no consensus on how trending ability, or trend analysis, should be performed. Therefore, we performed a literature review to identify articles published between 1997 and 2009 that compared methods of continuous CO measurement. Identified articles were grouped according to measurement technique and statistical methodology. Articles that analyzed trending ability were reviewed with the aim of finding an acceptable statistical method. Two hundred two articles were identified. The most popular methods were pulse contour (69 articles), Doppler (54), bioimpedance (38), and transpulmonary or continuous thermodilution (27). Forty-one articles addressed trending, and of these only 23 provided an in-depth analysis. Several common statistical themes were identified: time plots, regression analysis, Bland and Altman using change in CO (Delta CO), and the 4-quadrant plot, which used direction of change of Delta CO to determine the concordance. This plot was further refined by exclusion of data when values were small. Receiver operating characteristic curves were used to define the exclusion zone. In animal studies, a reliable reference standard such as an aortic flowprobe was frequently used, and regression or time plots could be used to show trending. Clinical studies were more problematic because data collection points were fewer (8-10 per subject). The consensus was to use the 4-quadrant plot with exclusion zones and apply concordance analysis. A concordance rate of >92% when using a 15% zone indicated good trending. A new method of presenting trend data (Delta CO) on a polar plot is proposed. Agreement was shown by the angle with the horizontal axis and Delta CO by the distance from the center. Trending can be assessed by the vertical limits of the data, similar to the Bland and Altman method. (Anesth Analg 2010;111:1180-92)
引用
收藏
页码:1180 / 1192
页数:13
相关论文
共 75 条
[1]   Equivalence of the bioimpedance and thermodilution methods in measuring cardiac output in hospitalized patients with advanced, decompensated chronic heart failure [J].
Albert, NM ;
Hail, MD ;
Li, JB ;
Young, JB .
AMERICAN JOURNAL OF CRITICAL CARE, 2004, 13 (06) :469-479
[2]   Comparison of transesophageal echocardiographic, Fick, and thermodilution cardiac output in critically ill patients [J].
Axler, O ;
Tousignant, C ;
Thompson, CR ;
DallavaSantucci, J ;
Phang, PT ;
Russell, JA ;
Walley, KR .
JOURNAL OF CRITICAL CARE, 1996, 11 (03) :109-116
[3]   Comparison of invasive and less-invasive techniques of cardiac output measurement under different haemodynamic conditions in a pig model [J].
Bajorat, J ;
Hofmockel, R ;
Vagts, DA ;
Janda, M ;
Pohl, B ;
Beck, C ;
Noeldge-Schomburg, G .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2006, 23 (01) :23-30
[4]   EVALUATION OF A TRANSIT-TIME SYSTEM FOR THE CHRONIC MEASUREMENT OF BLOOD-FLOW IN CONSCIOUS SHEEP [J].
BEDNARIK, JA ;
MAY, CN .
JOURNAL OF APPLIED PHYSIOLOGY, 1995, 78 (02) :524-530
[5]   Comparing different methods of cardiac output determination: a call for consensus [J].
Bein, B. ;
Renner, J. ;
Scholz, J. ;
Tonner, P. H. .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2006, 23 (08) :710-710
[6]   The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration [J].
Bein, Berthold ;
Meybohm, Patrick ;
Cavus, Erol ;
Renner, Jochen ;
Tonner, Peter H. ;
Steinfath, Markus ;
Scholz, Jens ;
Doerges, Volker .
ANESTHESIA AND ANALGESIA, 2007, 105 (01) :107-113
[7]   A NEW STROKE VOLUME EQUATION FOR THORACIC ELECTRICAL BIOIMPEDANCE - THEORY AND RATIONALE [J].
BERNSTEIN, DP .
CRITICAL CARE MEDICINE, 1986, 14 (10) :904-909
[8]   Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery [J].
Biancofiore, G. ;
Critchley, L. A. H. ;
Lee, A. ;
Bindi, L. ;
Bisa, M. ;
Esposito, M. ;
Meacci, L. ;
Mozzo, R. ;
DeSimone, P. ;
Urbani, L. ;
Filipponi, F. .
BRITISH JOURNAL OF ANAESTHESIA, 2009, 102 (01) :47-54
[9]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[10]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310