Six years' experience of using the BacT/ALERT system to screen all platelet concentrates, and additional testing of outdated platelet concentrates to estimate the frequency of false-negative results

被引:92
作者
Larsen, CP
Ezligini, F
Hermansen, NO
Kjeldsen-Kragh, J
机构
[1] Ullevaal Univ Hosp, Dept Immunol & Transfus Med, N-0407 Oslo, Norway
[2] Ullevaal Univ Hosp, Dept Microbiol, N-0407 Oslo, Norway
关键词
bacterial contamination; bacterial growth; blood culture; buffy-coat platelets; platelet apheresis; platelet concentrate;
D O I
10.1111/j.1423-0410.2005.00596.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives Approximately 1 in every 2000 units of platelets is contaminated with bacteria. The BacT/ALERT automated blood culture system can be used to screen platelet concentrates (PCs) for bacterial contamination. Materials and Methods Data were collected from May 1998 until May 2004. The number of PCs tested during this period was 36 896, most of which were produced from pools of four buffy-coats. On the day following blood collection or platelet apheresis, a 5-10 ml sample of the PC was aseptically transferred to a BacT/ALERT culture bottle for detection of aerobic bacteria. The sample was monitored for bacterial growth during the entire storage period of the PC (6.5 days). When a positive signal was generated, the culture bottle, the PC and the erythrocyte concentrates were tested for bacterial growth. In order to determine the frequency of false-negative BacT/ALERT signals, 1061 outdated PCs were tested during the period from May 2002 to May 2004. Results Eighty-eight positive signals were detected by the BacT/ALERT system, of which 12 were interpreted as truly positive. Fourteen signals were interpreted as truly false positive. Thirty-three signals were interpreted to be probably false positive. Two of 1061 outdated units tested positive, and Bacillus spp. and Staphylococcus epidermidis, respectively, were isolated from these PCs. Conclusions Between 0.03% and 0.12% of the PCs were contaminated with bacteria. BacT/ALERT is an efficient tool for monitoring PCs for bacterial contamination; however, it is important to realize that false-negative results may occur.
引用
收藏
页码:93 / 97
页数:5
相关论文
共 18 条
[1]   Hemovigilance network in France:: organization and analysis of immediate transfusion incident reports from 1994 to 1998 [J].
Andreu, G ;
Morel, P ;
Forestier, F ;
Debeir, J ;
Rebibo, D ;
Janvier, G ;
Hervé, P .
TRANSFUSION, 2002, 42 (10) :1356-1364
[2]   Experience with universal bacterial culturing to detect contamination of apheresis platelet units in a hospital transfusion service [J].
AuBuchon, JP ;
Cooper, LK ;
Leach, MF ;
Zuaro, DE ;
Schwartzman, JD .
TRANSFUSION, 2002, 42 (07) :855-861
[3]   Improving the bacteriological safety of platelet transfusions [J].
Blajchman, MA ;
Goldman, M ;
Baeza, F .
TRANSFUSION MEDICINE REVIEWS, 2004, 18 (01) :11-24
[4]  
BLAJCHMAN MA, 1996, BLOOD S, V88, P286
[5]   Evaluation of an automated culture system for detecting bacterial contamination of platelets: an analysis with 15 contaminating organisms [J].
Brecher, ME ;
Means, N ;
Jere, CS ;
Heath, D ;
Rothenberg, S ;
Stutzman, LC .
TRANSFUSION, 2001, 41 (04) :477-482
[6]   Evaluation of a new generation of culture bottle using an automated bacterial culture system for detecting nine common contaminating organisms found in platelet components [J].
Brecher, ME ;
Heath, DG ;
Hay, SN ;
Rothenberg, SJ ;
Stutzman, LC .
TRANSFUSION, 2002, 42 (06) :774-779
[7]   Monitoring of apheresis platelet bacterial contamination with an automated liquid culture system: a university experience [J].
Brecher, ME ;
Hay, SN ;
Rothenberg, SJ .
TRANSFUSION, 2003, 43 (07) :974-978
[8]  
Hogman CF, 1998, TRANSFUSION MED, V8, P1
[9]  
Katayama Toshio, 2003, Rinsho Ketsueki, V44, P381
[10]   Current status of microbial contamination of blood components: Summary of a conference [J].
Klein, HG ;
Dodd, RY ;
Ness, PM ;
Fratantoni, JA ;
Nemo, GJ .
TRANSFUSION, 1997, 37 (01) :95-101