Calibration effects for laser-induced breakdown spectroscopy of gaseous sample streams: Analyte response of gas-phase species versus solid-phase species

被引:67
作者
Hohreiter, V [1 ]
Hahn, DW [1 ]
机构
[1] Univ Florida, Dept Mech & Aerosp Engn, Gainesville, FL 32611 USA
关键词
D O I
10.1021/ac048587d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effects of analyte phase on the calibration response for laser-induced breakdown spectroscopy is investigated for a range of carbon species. Significant differences in the atomic emission signal from carbon were observed when comparing calibration streams of gas-phase and submicrometer-sized solid-phase carbon species. The resulting calibration curve slopes varied by a factor of 8 over a comparable range of atomic carbon concentrations for five different analyte sources, while the plasma electron density and temperature remained essentially constant. The current findings challenge a widely held assumption that complete dissociation of constituent species within a highly energetic laser-induced plasma results in independence of the analyte atomic emission signal on the analyte source. A physical model of the plasma-analyte interaction is proposed that provides a framework to account for the observed dependence on the physical state of the analyte.
引用
收藏
页码:1118 / 1124
页数:7
相关论文
共 33 条
[21]   AEROSOLS IN STRONG LASER-BEAMS [J].
LUSHNIKOV, AA ;
NEGIN, AE .
JOURNAL OF AEROSOL SCIENCE, 1993, 24 (06) :707-&
[22]   Detection of bacteria by time-resolved laser-induced breakdown spectroscopy [J].
Morel, S ;
Leone, N ;
Adam, P ;
Amouroux, J .
APPLIED OPTICS, 2003, 42 (30) :6184-6191
[23]   Time-resolved microscopic imaging of the laser-induced forward transfer process [J].
Nakata, Y ;
Okada, T .
APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1999, 69 (Suppl 1) :S275-S278
[24]   On-line and in-situ detection of lead aerosols by plasma-spectroscopy and laser-excited atomic fluorescence spectroscopy [J].
Neuhauser, RE ;
Panne, U ;
Niessner, R ;
Petrucci, GA ;
Cavalli, P ;
Omenetto, N .
ANALYTICA CHIMICA ACTA, 1997, 346 (01) :37-48
[25]   On-line monitoring of chromium aerosols in industrial exhaust streams by laser-induced plasma spectroscopy (LIPS) [J].
Neuhauser, RE ;
Panne, U ;
Niessner, R ;
Wilbring, P .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 364 (08) :720-726
[26]   Analysis of sulfuric acid aerosols by laser-induced breakdown spectroscopy and laser-induced photofragmentation [J].
Núñez, MH ;
Cavalli, P ;
Petrucci, G ;
Omenetto, N .
APPLIED SPECTROSCOPY, 2000, 54 (12) :1805-1816
[27]   TIME-RESOLVED LASER-INDUCED BREAKDOWN SPECTROMETRY OF AEROSOLS [J].
RADZIEMSKI, LJ ;
LOREE, TR ;
CREMERS, DA ;
HOFFMAN, NM .
ANALYTICAL CHEMISTRY, 1983, 55 (08) :1246-1252
[28]   Fundamentals and applications of laser-induced breakdown spectroscopy [J].
Rusak, DA ;
Castle, BC ;
Smith, BW ;
Winefordner, JD .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1997, 27 (04) :257-290
[29]   Laser-induced breakdown spectroscopy of bacterial spores, molds, pollens, and protein: initial studies of discrimination potential [J].
Samuels, AC ;
DeLucia, FC ;
McNesby, KL ;
Miziolek, AW .
APPLIED OPTICS, 2003, 42 (30) :6205-6209
[30]   Molecular dynamics simulation of the laser disintegration of aerosol particles [J].
Schoolcraft, TA ;
Constable, GS ;
Zhigilei, LV ;
Garrison, BJ .
ANALYTICAL CHEMISTRY, 2000, 72 (21) :5143-5150