Noninvasive MR imaging-guided focal opening of the blood-brain barrier in rabbits

被引:1133
作者
Hynynen, K [1 ]
McDannold, N [1 ]
Vykhodtseva, N [1 ]
Jolesz, FA [1 ]
机构
[1] Harvard Univ, Brigham & Womens Hosp, Sch Med, Dept Radiol, Boston, MA 02115 USA
关键词
blood-brain barrier; experimental study; magnetic resonance (MR); guidance; temperature monitoring; ultrasound; (US); therapeutic;
D O I
10.1148/radiol.2202001804
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine if focused ultrasound beams can be used to locally open the blood-brain barrier without damage to surrounding brain tissue and if magnetic resonance (MR) imaging can be used to monitor this procedure. MATERIALS AND METHODS: The brains of 18 rabbits were sonicated (pulsed sonication) in four to six locations, with temporal peak acoustic power ranging from 0.2 to 11.5 W. Prior to each sonication, a bolus of ultrasonographic (US) contrast agent was injected into the ear vein of the rabbit. A series of fast or spoiled gradient-echo MR images were obtained during the sonications to monitor the temperature elevation and potential tissue changes. Contrast material-enhanced MR images obtained minutes after sonications and repeated 1-48 hours later were used to depict blood-brain barrier opening. Whole brain histologic evaluation was performed. RESULTS: Opening of the blood-brain barrier was confirmed with detection of MR imaging contrast agent at the targeted locations. The lowest power levels used produced blood-brain barrier opening without damage to the surrounding neurons. Contrast enhancement correlated with the focal signal intensity changes in the magnitude fast spoiled gradient-echo MR images. CONCLUSION: The blood-brain barrier can be consistently opened with focused ultrasound exposures in the presence of a US contrast agent. MR imaging signal intensity changes may be useful in the detection of blood-brain barrier opening during sonication.
引用
收藏
页码:640 / 646
页数:7
相关论文
共 27 条
[1]   Transporting therapeutics across the blood-brain barrier [J].
Abbott, NJ ;
Romero, IA .
MOLECULAR MEDICINE TODAY, 1996, 2 (03) :106-113
[2]   In vivo target-specific delivery of macromolecular agents with MR-guided focused ultrasound [J].
Bednarski, MD ;
Lee, JW ;
Callstrom, MR ;
Li, KCP .
RADIOLOGY, 1997, 204 (01) :263-268
[3]  
CARWASH NH, 1999, CRIT REV ONCOGEN, V10, P261
[4]   Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam [J].
Chung, AH ;
Hynynen, K ;
Colucci, V ;
Oshio, K ;
Cline, HE ;
Jolesz, FA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) :745-752
[5]   A hemisphere array for non-invasive ultrasound brain therapy and surgery [J].
Clement, GT ;
Sun, J ;
Giesecke, T ;
Hynynen, K .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12) :3707-3719
[6]   Hemolysis in vivo from exposure to pulsed ultrasound [J].
Dalecki, D ;
Raeman, CH ;
Child, SZ ;
Cox, C ;
Francis, CW ;
Meltzer, RS ;
Carstensen, EL .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (02) :307-313
[7]   The influence of contrast agents on hemorrhage produced by lithotripter fields [J].
Dalecki, D ;
Raeman, CH ;
Child, SZ ;
Penney, DP ;
Mayer, R ;
Carstensen, EL .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (09) :1435-1439
[8]   Design and evaluation of a feedback based phased array system for ultrasound surgery [J].
Daum, DR ;
Buchanan, MT ;
Fjield, T ;
Hynynen, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (02) :431-438
[9]   Histological evidence of fetal pig neural cell survival after transplantation into a patient with Parkinson's disease [J].
Deacon, T ;
Schumacher, J ;
Dinsmore, J ;
Thomas, C ;
Palmer, P ;
Kott, S ;
Edge, A ;
Penney, D ;
Kassissieh, S ;
Dempsey, P ;
Isacson, O .
NATURE MEDICINE, 1997, 3 (03) :350-353
[10]  
Doolittle ND, 2000, CANCER, V88, P637, DOI 10.1002/(SICI)1097-0142(20000201)88:3<637::AID-CNCR22>3.0.CO