Preparation of nano-sized Pb2+ imprinted polymer and its application as the chemical interface of an electrochemical sensor for toxic lead determination in different real samples

被引:93
作者
Alizadeh, Taher [1 ]
Amjadi, Somaye [1 ]
机构
[1] Univ Mohaghegh Ardabili, Fac Sci, Dept Appl Chem, Ardebil 5619911367, Iran
关键词
Pb2+; Imprinted polymer; Differential pulse voltammetry; Electrochemical sensor; ANODIC-STRIPPING VOLTAMMETRY; SOLID-PHASE EXTRACTION; CARBON-PASTE ELECTRODE; BISMUTH-FILM ELECTRODES; TRACE DETERMINATION; SPECTROPHOTOMETRIC DETERMINATION; PLATED CARBON; HEAVY-METALS; WATER; CADMIUM;
D O I
10.1016/j.jhazmat.2011.03.067
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In this work, a new nano-structured ion imprinted polymer (IIP) was synthesized by copolymerization of methacrylic acid-Pb2+ complex and ethylene glycol dimethacrylate according to the precipitation polymerization. Methacrylic acid acted as both functional monomer and complexing agent to create selective coordination sites in a cross-linked polymer. A carbon paste electrode modified with IIP-nanoparticles was used for fabrication of a Pb2+ sensitive electrode. Differential pulse stripping voltammetry method was applied as the determination technique, after open circuit sorption of Pb2+ on the electrode and its reduction to metallic form. The IIP modified electrode showed a considerably higher response, compared to the electrode embedded with non-imprinted polymer (NIP). This indicated that the suitable recognition sites were created in the IIP structure in the polymerization stage. Various factors, effective on the response behavior of the electrode, were investigated and optimized. The introduced sensor showed a linear range of 1.0 x 10(-9) to 8.1 x 10(-7) M and detection limit of 6.0 x 10(-19) M (S/N = 3). The sensor was successfully applied for the trace lead determination in different samples. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:451 / 459
页数:9
相关论文
共 85 条
[71]   Electrochemical determination of lead and glutathione in a plant cell culture [J].
Vacek, J ;
Petrek, J ;
Kizek, R ;
Havel, L ;
Klejdus, B ;
Trnková, L ;
Jelen, F .
BIOELECTROCHEMISTRY, 2004, 63 (1-2) :347-351
[72]   DRUG ASSAY USING ANTIBODY MIMICS MADE BY MOLECULAR IMPRINTING [J].
VLATAKIS, G ;
ANDERSSON, LI ;
MULLER, R ;
MOSBACH, K .
NATURE, 1993, 361 (6413) :645-647
[73]   Processes during anodic stripping voltammetry determination of lead in the presence of copper on a solid electrode modified with 2,2′-bipyridyl in polyaniline [J].
Wagner, K ;
Strojek, JW ;
Koziel, K .
ANALYTICA CHIMICA ACTA, 2001, 447 (1-2) :11-21
[74]   Bismuth-coated carbon electrodes for anodic stripping voltammetry [J].
Wang, J ;
Lu, JM ;
Hocevar, SB ;
Farias, PAM ;
Ogorevc, B .
ANALYTICAL CHEMISTRY, 2000, 72 (14) :3218-3222
[75]   Stripping voltammetric determination of Pb(II) and Cd(II) based on the multiwalled carbon nanotubes-Nafion-bismuth modified glassy carbon electrodes [J].
Wang, Na ;
Dong, Xiandui .
ANALYTICAL LETTERS, 2008, 41 (07) :1267-1278
[76]   Molecular imprinting and solid phase extraction of flavonoid compounds [J].
Weiss, R ;
Molinelli, A ;
Jakusch, M ;
Mizaikoff, B .
BIOSEPARATION, 2001, 10 (06) :379-387
[77]   Lead(II) ion selective electrodes with PVC membranes based on two bis-thioureas as ionophores: 1,3-bis(N′-benzoylthioureido)benzene and 1,3-bis(N′-furoylthioureido)benzene [J].
Wilson, Deivy ;
de los Angeles Arada, Maria ;
Alegret, Salvador ;
del Valle, Manel .
JOURNAL OF HAZARDOUS MATERIALS, 2010, 181 (1-3) :140-146
[78]   Simultaneous measurement of Pb, Cd and Zn using differential pulse anodic stripping voltammetry at a bismuth/poly(p-aminobenzene sulfonic acid) film electrode [J].
Wu, Yun ;
Li, Nian Bing ;
Luo, Hong Qun .
SENSORS AND ACTUATORS B-CHEMICAL, 2008, 133 (02) :677-681
[79]   Direct extraction of specific pharmacophoric flavonoids from gingko leaves using a molecularly imprinted polymer for quercetin [J].
Xie, JC ;
Zhu, LL ;
Luo, HP ;
Zhou, L ;
Li, CX ;
Xu, XJ .
JOURNAL OF CHROMATOGRAPHY A, 2001, 934 (1-2) :1-11
[80]   Screen-printed electrodes modified with functionalized mesoporous silica for voltammetric analysis of toxic metal ions [J].
Yantasee, W ;
Deibler, LA ;
Fryxell, GE ;
Timchalk, C ;
Lin, YH .