Stroke volume and cardiac output evaluation by electrical cardiometry: accuracy and reference nomograms in hemodynamically stable preterm neonates

被引:53
作者
Boet, A. [1 ,2 ]
Jourdain, G. [2 ]
Demontoux, S. [1 ]
De Luca, D. [2 ,3 ]
机构
[1] South Paris Univ Hosp, Marie Lannelongue Surg Ctr, Cardiac ICU, Paris, France
[2] South Paris Univ Hosp, AP HP, Div Paediat & Neonatal Crit Care, A Beclere Med Ctr, Paris, France
[3] Univ Cattolica Sacro Cuore, Inst Anaesthesiol & Crit Care, Rome, Italy
关键词
VELOCIMETRY; CHILDREN; ECHOCARDIOGRAPHY; THERMODILUTION; REGRESSION; INFANTS;
D O I
10.1038/jp.2016.65
中图分类号
R71 [妇产科学];
学科分类号
100211 [妇产科学];
摘要
OBJECTIVE: To investigate the accuracy of electrical cardiometry (EC) to measure stroke volume (SV) and cardiac output (CO) and to provide gestational age (GA) and birth weight (BW)-based reference data for SV and CO in hemodynamically stable preterm neonates. STUDY DESIGN: Prospective observational blinded study. Paired measurements of SV and CO on stable preterm infants without any hemodynamic compromise were carried out using EC (SVEC) and echocardiography (SVECHO). RESULTS: Seventy-nine preterm neonates (mean GA: 31 +/- 3.2 weeks) were enrolled. A good correlation was found for SV (r = 0.743; P < 0.0001) and CO (r = 0.7; P < 0.0001) measured by EC and echocardiography. These correlations remained significant after adjusting for GA, patent ductus arteriosus and type of respiratory support (SV: St.beta = 0.48, P < 0.0001 and CO: st.beta = 0.69, P < 0.0001). Mean biases (and variabilities) were -1.1 (from 0.7 to -2.9) ml and -0.21 (from 0.15 to -0.55) I min(-1) for SV and CO, respectively. Local regression shows a tendency for EC to overestimate SV and CO especially at higher values (at about >2 ml and >0.41 min(-1), respectively). Coefficient of variation of SV was 48.9% and 52%, for EC and echocardiography. SV and CO rose with increasing GA and BW following an exponential equation (R-2>0.8). CONCLUSION: Measuring SV and CO with EC in hemodynamically stable preterm infants shows good correlation and variability similar to that of echocardiography. A trend to overestimation exists at highest values, but it is unlikely to be clinically significant. Reference GA and BW-based nomograms for SV and CO are provided.
引用
收藏
页码:748 / 752
页数:5
相关论文
共 25 条
[1]
Electrical velocimetry demonstrates the increase in cardiac output and decrease in systemic vascular resistance accompanying cesarean delivery and oxytocin administration [J].
Archer, Thomas L. ;
Conrad, Benjamin E. ;
Suresh, Preetham ;
Tarsa, Maryam .
JOURNAL OF CLINICAL ANESTHESIA, 2012, 24 (01) :79-82
[2]
Stroke volume equation for impedance cardiography [J].
Bernstein, DP ;
Lemmens, HJM .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (04) :443-450
[3]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]
Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study [J].
Blohm, Martin Ernst ;
Obrecht, Denise ;
Hartwich, Jana ;
Mueller, Goetz Christoph ;
Kersten, Jan Felix ;
Weil, Jochen ;
Singer, Dominique .
CRITICAL CARE, 2014, 18 (06)
[5]
Cameron AC, 1997, J ECONOMETRICS, V77, P329
[6]
Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies - with specific reference to the measurement of cardiac output [J].
Cecconi, Maurizio ;
Rhodes, Andrew ;
Poloniecki, Jan ;
Della Rocca, Giorgio ;
Grounds, R. Michael .
CRITICAL CARE, 2009, 13 (01)
[8]
Attempt to improve transcutaneous bilirubinometry: a double-blind study of Medick BiliMed versus Respironics BiliCheck [J].
De Luca, D. ;
Zecca, E. ;
Corsello, M. ;
Tiberi, E. ;
Semeraro, C. ;
Romagnoli, C. .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2008, 93 (02) :F135-F139
[9]
Goldstein Brahm, 2005, Pediatr Crit Care Med, V6, P2
[10]
Non-invasive cardiac output measure end in low and very low birth weight infants: a method comparison [J].
Grollmuss, Oswin ;
Gonzalez, Patricia .
FRONTIERS IN PEDIATRICS, 2014, 2