Dosimetry with radiochromic films: a document scanner technique, neutron response, applications

被引:28
作者
Bazioglou, M [1 ]
Kalef-Ezra, J [1 ]
机构
[1] Univ Ioannina, Sch Med, Med Phys Lab, GR-45110 Ioannina, Greece
关键词
radiochromic films; neutrons; dosimetry; document scanner;
D O I
10.1016/S0969-8043(01)00060-4
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The possibility of reading radiochromic films by the employment of a document scanner and associated software was explored. Dosimetric characteristics, such as the sensitivity of the technique, reproducibility in dose assessment, minimum detectable dose, film uniformity, OD variation with time, variation of OD due to multiple scans, polarity of the analyzing light, fast neutron response and thermal neutron yield were investigated. The technique was used to measure dose distributions in stereotactic radiotherapy, orthovoltage treatment and neutron capture therapy. (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:339 / 345
页数:7
相关论文
共 25 条
[1]   Assessment of large single-fraction, low-energy X-ray dose with radiochromic film [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN ;
Metcalfe, PE .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (04) :1071-1075
[2]   HIGH-SENSITIVITY GAFCHROMIC FILM DOSIMETRY FOR I-125 SEED [J].
CHIUTSAO, ST ;
DELAZERDA, A ;
LIN, J ;
KIM, JH .
MEDICAL PHYSICS, 1994, 21 (05) :651-657
[3]   The effects of boron neutron capture irradiation on oral mucosa: evaluation using a rat tongue model [J].
Coderre, JA ;
Morris, GM ;
Kalef-Ezra, J ;
Micca, PL ;
Ma, R ;
Youngs, K ;
Gordon, CR .
RADIATION RESEARCH, 1999, 152 (02) :113-118
[4]   A novel 675.2 nm diode laser densitometer for use with GafChromic films [J].
Ertl, A ;
Kellermann, PO ;
Zehetmayer, M ;
Schöggl, A ;
Kindl, P ;
Maitz, AH .
MEDICAL PHYSICS, 1999, 26 (05) :834-838
[5]  
FAIRCHILD RG, 1988, P 1988 INT REACT PHY, V2, P423
[6]   A comparison of dosimetry techniques in stereotactic radiosurgery [J].
Heydarian, M ;
Hoban, PW ;
Beddoe, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :93-110
[7]  
KALEFEZRA J, 1999, PHYS MED, V15, P222
[8]  
Kappas C, 1997, PHYS MEDICA, V13, P123
[9]   GafChromic MD-55: Investigated as a precision dosimeter [J].
Klassen, NV ;
van der Zwan, L ;
Cygler, J .
MEDICAL PHYSICS, 1997, 24 (12) :1924-1934
[10]   IMPROVED APPARATUS FOR NEUTRON-CAPTURE THERAPY OF RAT-BRAIN TUMORS [J].
LIU, HYB ;
JOEL, DD ;
SLATKIN, DN ;
CODERRE, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (05) :1167-1173