Autonomous microfluidic multi-channel chip for real-time PCR with integrated liquid handling

被引:48
作者
Frey, Olivier
Bonneick, Sonja
Hierlemann, Andreas
Lichtenberg, Jan
机构
[1] Univ Neuchatel, Inst Microtechnol, Sensors Actuators & Microsyst Lab, CH-2000 Neuchatel, Switzerland
[2] ETH, Phys Elect Lab, CH-8093 Zurich, Switzerland
关键词
PCR; Microchip PCR; sample shuttling; PDMS; diffuser nozzle structures;
D O I
10.1007/s10544-007-9080-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We report on a novel, polymer-based, multi-channel device for polymerase chain reaction that combines, for the first time, rapid sample processing in less than 5 min with high throughput at low costs. This is achieved by sample shuttling, during which submicroliter sample plugs (similar to 100 nl) are oscillated rapidly over three constant-temperature zones by pneumatic actuation with integrated system. The accuracy and the speed of the liquid handling have been significantly increased, while the design of the device can be kept very simple and allows for mass production using conventional low-cost polymer fabrication processes. Massive parallelization can lead to a throughput up to 100 samples in 10 min including the preparation time. The amplification can be optically monitored by means of online fluorescence detection. Successful real-time PCR and the determination of the threshold cycle, C-t, using the developed device were demonstrated with plasmid DNA in a fluorescent real-time format.
引用
收藏
页码:711 / 718
页数:8
相关论文
共 15 条
[1]   Miniaturised nucleic acid analysis [J].
Auroux, PA ;
Koc, Y ;
deMello, A ;
Manz, A ;
Day, PJR .
LAB ON A CHIP, 2004, 4 (06) :534-546
[2]  
Auroux PA, 2004, NSTI NANOTECH 2004, VOL 1, TECHNICAL PROCEEDINGS, P67
[3]   An integrated nanoliter DNA analysis device [J].
Burns, MA ;
Johnson, BN ;
Brahmasandra, SN ;
Handique, K ;
Webster, JR ;
Krishnan, M ;
Sammarco, TS ;
Man, PM ;
Jones, D ;
Heldsinger, D ;
Mastrangelo, CH ;
Burke, DT .
SCIENCE, 1998, 282 (5388) :484-487
[4]   Polymerase chain reaction in polymeric microchips:: DNA amplification in less than 240 seconds [J].
Giordano, BC ;
Ferrance, J ;
Swedberg, S ;
Hühmer, AFR ;
Landers, JP .
ANALYTICAL BIOCHEMISTRY, 2001, 291 (01) :124-132
[5]   Chemical amplification: Continuous-flow PCR on a chip [J].
Kopp, MU ;
de Mello, AJ ;
Manz, A .
SCIENCE, 1998, 280 (5366) :1046-1048
[6]   Fully integrated PCR-capillary electrophoresis microsystem for DNA analysis [J].
Lagally, ET ;
Emrich, CA ;
Mathies, RA .
LAB ON A CHIP, 2001, 1 (02) :102-107
[7]   Miniaturization of polymerase chain reaction [J].
Lee, JY ;
Kim, JJ ;
Park, TH .
BIOTECHNOLOGY AND BIOPROCESS ENGINEERING, 2003, 8 (04) :213-220
[8]  
Liu J, 2002, ELECTROPHORESIS, V23, P1531, DOI 10.1002/1522-2683(200205)23:10<1531::AID-ELPS1531>3.0.CO
[9]  
2-D
[10]  
Northrup MA., 1993, TRANSDUCERS, P924