Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers - bone mineral density and body composition studies

被引:58
作者
Oldroyd, B [1 ]
Smith, AH [1 ]
Truscott, JG [1 ]
机构
[1] Univ Leeds, Ctr Bone & Body Composit Res, Leeds LS2 9JT, W Yorkshire, England
关键词
dual-energy X-ray absorptiometers; cross-calibration; pencil beam; fan beam; body composition; bone mineral;
D O I
10.1038/sj.ejcn.1601633
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Objective: In vitro and in vivo comparisons of bone mineral density (BMD) and body composition between GE/Lunar pencil (DPXL) and fan-beam (PRODIGY) absorptiometers. Design: Comparison of BMD, bone mineral content (BMC) and area of lumbar spine (L2-L4), femoral neck and total body. Total body composition compartments tissue (TBTissue), fat (TBF), lean tissue (TBLean) and %TBF were also compared. Setting: Centre for Bone and Body Composition Research, University of Leeds. Phantoms/subjects: A range of spine phantoms, a variable composition phantom (VCP) and total body phantom. A total of 72 subjects were included for the in vivo study. Results: In vitro: A small significant underestimation of BMD by the Prodigy compared to the DPXL ranging from 0.7 to 2% (p < 0.05-0.001) for the spine phantoms. The Prodigy underestimated the VCP %Fat. Although the Prodigy underestimated phantom TBBMD by 1.1 +/- 1.0%, TBBMC and area were reduced by 8.2 +/- 1.4 and 7.3 +/- 1.0%, respectively. The Prodigy overestimated TBTissue 1508 g (2.2%), TBLean 588 g (1.2%), TBF 919 g (4.8%) and %TBF (0.8%). In vivo: BMD cross-calibration was only required in the femoral neck, DPXLBMD = 0.08+0.906*PRODIGY(BMD). The Prodigy had higher estimates for TBTissue 1360 g (2.3%), TBLean 840 g (2.0%), TBF 519 g (3.4%), TBBMC 32.8 g (1.3%) and %TBF (0.3%). Cross-calibration equations were required for TBTissue(DPXL) = -1158+0.997*TBTissue(PRODIGY) and TBBMCDPXL = 89.7+0.949*TBBMCPRODIGY. Conclusions: Small differences between the two absorptiometers for both BMD and body composition can be made compatible by use of cross-calibration equations and factors. The discrepancy in body composition compartments requires further research.
引用
收藏
页码:977 / 987
页数:11
相关论文
共 36 条
[1]   CROSS CALIBRATION OF QDR-2000 AND QDR-1000 DUAL-ENERGY X-RAY DENSITOMETERS FOR BONE-MINERAL AND SOFT-TISSUE MEASUREMENTS [J].
ABRAHAMSEN, B ;
GRAM, J ;
HANSEN, TB ;
BECKNIELSEN, H .
BONE, 1995, 16 (03) :385-390
[2]   Comparison of two Hologic DXA systems (QDR 1000 and QDR 4500/A) [J].
Barthe, N ;
Braillon, P ;
Ducassou, D ;
BasseCathalinat, B .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (835) :728-739
[3]  
Blake Glen M., 1998, EVALUATION OSTEOPORO
[4]   DUAL X-RAY ABSORPTIOMETRY - A COMPARISON BETWEEN FAN-BEAM AND PENCIL BEAM SCANS [J].
BLAKE, GM ;
PARKER, JC ;
BUXTON, FMA ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) :902-906
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   Cross-calibration of a fan-beam X-ray densitometer with a pencil-beam system [J].
Bouyoucef, SE ;
Cullum, ID ;
Jell, P .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (822) :522-531
[7]   Evaluation of a new body composition phantom for quality control and cross-calibration of DXA devices [J].
Diessel, E ;
Fuerst, T ;
Njeh, CF ;
Tylavsky, F ;
Cauley, J ;
Dockrell, M ;
Genant, HK .
JOURNAL OF APPLIED PHYSIOLOGY, 2000, 89 (02) :599-605
[8]   A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurement [J].
Economos, CD ;
Nelson, ME ;
Fiatarone, MA ;
Dallal, GE ;
Heymsfield, SB ;
Wang, J ;
Yasumara, S ;
Ma, R ;
Vaswani, AN ;
RussellAulet, M ;
Pierson, RN .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 1997, 51 (05) :312-317
[9]   SWITCHING FROM DXA PENCIL-BEAM TO FAN-BEAM .2. STUDIES IN-VIVO [J].
EIKEN, P ;
KOLTHOFF, N ;
BARENHOLDT, O ;
HERMANSEN, F ;
NIELSEN, SP .
BONE, 1994, 15 (06) :671-676
[10]   SWITCHING FROM DXA PENCIL-BEAM TO FAN-BEAM .1. STUDIES IN-VITRO AT 4 CENTERS [J].
EIKEN, P ;
BARENHOLDT, O ;
JENSEN, LB ;
GRAM, J ;
NIELSEN, SP .
BONE, 1994, 15 (06) :667-670