A valve-less diffuser micropump for microfluidic analytical systems

被引:185
作者
Andersson, H
van der Wijngaart, W
Nilsson, P
Enoksson, P
Stemme, G
机构
[1] Royal Inst Technol, Instrumentat Lab, Dept Signals Sensors & Syst, S-10044 Stockholm, Sweden
[2] Royal Inst Technol, Instrumentat Lab, Dept Biotechnol, S-10044 Stockholm, Sweden
关键词
valve-less micropump; highly viscous liquids; bead handling; living cells;
D O I
10.1016/S0925-4005(00)00644-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The suitability of valve-less micropumps in biochemistry has been shown. Fluids encountered in various biochemical methods that are problematic for other micropumps have been pumped with good performance. The pump is fabricated as a silicon-glass stack with a new process involving three subsequent deep reactive ion etching steps. Some of the main advantages of the valve-less diffuser pump are the absence of moving parts (excluding the pump diaphragm), the uncomplicated planar design, and high pump performance in terms of pressure head and flow rare. In addition, the micropump is self-priming and insensitive to particles and bubbles present in the pumped media. The results show that the valve-less micropump successfully pumps fluids within the viscosity range of 0.001-0.9 N s/m(2). The micropump is not sensitive to the density, ionic strength, or pH of the pumped media. Effective pumping of solutions containing beads of different sizes was also demonstrated. Living cells were pumped without inducing cell damage and no cell adhesion within the pump chamber was found. No valve-less micropump has previously been reported to pump such a wide variety of fluids. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 15 条
[1]   MICROFABRICATED ELECTROHYDRODYNAMIC PUMPS [J].
BART, SF ;
TAVROW, LS ;
MEHREGANY, M ;
LANG, JH .
SENSORS AND ACTUATORS A-PHYSICAL, 1990, 21 (1-3) :193-197
[2]   Dielectrophoretic sorting of particles and cells in a microsystem [J].
Fiedler, S ;
Shirley, SG ;
Schnelle, T ;
Fuhr, G .
ANALYTICAL CHEMISTRY, 1998, 70 (09) :1909-1915
[3]   Microfluidics - a review [J].
Gravesen, Peter ;
Branebjerg, Jens ;
Jensen, Ole Sondergard .
Journal of Micromechanics and Microengineering, 1993, 3 (04) :168-182
[4]   Transport, manipulation, and reaction of biological cells on-chip using electrokinetic effects [J].
Li, PCH ;
Harrison, DJ .
ANALYTICAL CHEMISTRY, 1997, 69 (08) :1564-1568
[5]   ELECTROOSMOTIC PUMPING AND ELECTROPHORETIC SEPARATIONS FOR MINIATURIZED CHEMICAL-ANALYSIS SYSTEMS [J].
MANZ, A ;
EFFENHAUSER, CS ;
BURGGRAF, N ;
HARRISON, DJ ;
SEILER, K ;
FLURI, K .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1994, 4 (04) :257-265
[6]   Microfabricated devices for genetic diagnostics [J].
Mastrangelo, CH ;
Burns, MA ;
Burke, DT .
PROCEEDINGS OF THE IEEE, 1998, 86 (08) :1769-1787
[7]   Micromachined flat-walled valveless diffuser pumps [J].
Olsson, A ;
Enoksson, P ;
Stemme, G ;
Stemme, E .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 1997, 6 (02) :161-166
[8]  
QIU X, 1997, P 1997 INT C SOL STA, P923
[9]   Micropump for viscous liquids and muds [J].
Schwesinger, N ;
Bechtel, S .
MICROFLUIDIC DEVICES AND SYSTEMS, 1998, 3515 :40-45
[10]   MICROFLOW DEVICES AND SYSTEMS [J].
SHOJI, S ;
ESASHI, M .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1994, 4 (04) :157-171