摘要
A method was developed to determine nine N-nitrosamines in wastewater on, the basis of solid-phase extraction and liquid chromatography mass spectrometry using a linear ion trap-orbitrap hybrid instrument at high mass resolution. Analytes and five deuterated internal standards were preconcentrated by solid-phase extraction. Positive electrospray ionization resulted in protonated molecular ions of all nitrosamines. One to three product ions were formed by collision-induced dissociation or higher energy C-trap dissociation. The signal intensity of the product ions differed up to a factor of 3 between the two techniques. The molecular ions were usually used for quantification, because of the better sensitivity, and the product ions for confirmation. An actual mass resolving power of 25 000-40 000 ensured a sufficient selectivity to distinguish all molecular and product ions from interfering background ions. Only for N-nitrosomorpholine was a coeluting isobaric molecular ion detected in wastewater samples, which, however, formed different product ions, The mass accuracy was between - 12 ppm at m/z 55 and 0 ppm at m/z 205 and did not change for more than 5 ppm over a sample sequence of 20 h analysis time. The optimized method allowed quantifying nine N-nitrosamines in drinking water and wastewater samples down to method detection limits of 0.3-3.9 ng/L at instrumental detection limits of 2-14 pg on column. Recoveries over the whole method were between 75 and 125% for six compounds, but considerably lower for three compounds, probably due to strong matrix effects causing a signal suppression of up to 95% in wastewater samples. N-Nitrosodimethylamine and N-nitrosomorpholine were the most abundant compounds (3-22 ng/L) in samples from two wastewater treatment plants, another four nitrosamines (N-nitrosopyrrolidone, -piperidine, -diethylamine, and -dibutylamine) were also detected. Our study demonstrates that the LTQ Orbitrap is a powerful instrument to quantify low molecular weight compounds at the picogram level in complex matrixes with both a high sensitivity and selectivity.
引用
收藏
页码:834 / 842
页数:9
相关论文
共 27 条
[1]  
[Anonymous], 1991, ENV SAMPLING ANAL PR
[2]   Detecting N-nitrosamines in drinking water at nanogram per liter levels using ammonia positive chemical ionization [J].
Charrois, JWA ;
Arend, MW ;
Froese, KL ;
Hrudey, SE .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2004, 38 (18) :4835-4841
[3]  
Cheng RC, 2006, J AM WATER WORKS ASS, V98, P82
[4]  
*GLOB WAT RES COLL, 2007, REP GWRC RES STRAT W
[5]   The Orbitrap: a new mass spectrometer [J].
Hu, QZ ;
Noll, RJ ;
Li, HY ;
Makarov, A ;
Hardman, M ;
Cooks, RG .
JOURNAL OF MASS SPECTROMETRY, 2005, 40 (04) :430-443
[6]   Analysis of tobacco-specific N-nitrosamines in snuff by ethyl acetate extraction and liquid chromatography-tandem mass spectrometry [J].
Jansson, C ;
Paccou, A ;
Österdahl, BG .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1008 (02) :135-143
[7]  
LALLJIE SPD, 1998, ADV MASS SPECTROMETR, V14
[8]   Metabolite identification by data-dependent accurate mass spectrometric analysis at resolving power of 60,000 in external calibration mode using an LTQ/Orbitrap [J].
Lim, Heng-Keang ;
Chen, Jie ;
Sensenhauser, Carlo ;
Cook, Kevin ;
Subrahmanyam, Vangala .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (12) :1821-1832
[9]   Performance evaluation of a hybrid linear ion trap/orbitrap mass spectrometer [J].
Makarov, A ;
Denisov, E ;
Kholomeev, A ;
Baischun, W ;
Lange, O ;
Strupat, K ;
Horning, S .
ANALYTICAL CHEMISTRY, 2006, 78 (07) :2113-2120
[10]   Dynamic range of mass accuracy in LTQ Orbitrap hybrid mass spectrometer [J].
Makarov, Alexander ;
Denisov, Eduard ;
Lange, Oliver ;
Horning, Stevan .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2006, 17 (07) :977-982