Commissioning of a micro multi-leaf collimator and planning system for stereotactic radiosurgery

被引:137
作者
Cosgrove, VP [1 ]
Jahn, U [1 ]
Pfaender, M [1 ]
Bauer, S [1 ]
Budach, V [1 ]
Wurm, RE [1 ]
机构
[1] Free Univ Berlin, Klinikum Charite, Klin Strathlentherapie, D-10117 Berlin, Germany
关键词
micro multi-leaf collimator; MLC; conformal therapy; dosimetry; radiosurgery planning;
D O I
10.1016/S0167-8140(99)00020-1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A computer controlled micro multi-leaf collimator, m3 mMLC, has been commissioned for conformal, fixed-held radiosurgery applications. Measurements were made to characterise the basic dosimetric properties of the m3, such as leaf transmission, leakage and beam penumbra. In addition, the geometric and dosimetric accuracy of the m3 was verified when used in conjunction with a BrainSCAN v3.5 stereotactic planning system. Materials and methods: The m3 was detachably mounted to a Varian Clinac 2100C accelerator delivering 6 MV X-rays. Leaf transmission, leakage, penumbra and multiple, conformal fixed field dose distributions were measured using calibrated film in solid water. Beam data were collected using a diamond detector in a scanning water tank and planned dose distributions were verified using LIF TLDs and film. A small, shaped phantom was also constructed to confirm field shaping accuracy using portal images. Results: Mean transmission through the closed multi-leaves was 1.9 +/- 0.1% and leakage between leaves was 2.8 +/- 0.15%. Between opposing leaves abutting along the central beam-axis transmission was similar to 15 +/- 3%, but was reduced to a mean of 4.5 +/- 0.6% by moving the abutmen position 4.5 cm off-axis. Beam penumbrae were effectively constant as a function of increasing square field size and asymmetric fields and was seen to vary non-linearly when shaped to diagonal, straight edges. TMR, OAR and relative output beam data measurements of circular m3 fields were comparable to conventional, circular stereotactic collimators. Multiple, conformal field dose distributions were calculated with good spatial and dosimetric accuracy, with the planned 90% isodose curves agreeing with measurements to within 1-2 mm and to +/- 3% at isocentre. Portal films agreed with planned beams eye-view field shaping to within 1 mm. Conclusions: The m3 micro multi-leaf collimator is a stable, high precision field-shaping device suitable for small-field, radiosurgery applications. Dose distributions can be accurately calculated by a planning system using only a few beam data parameters. (C) 1999 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:325 / 336
页数:12
相关论文
共 45 条
[1]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[2]   STATIC-FIELD CONFORMAL STEREOTAXIC RADIOSURGERY - PHYSICAL TECHNIQUES [J].
BOURLAND, JD ;
MCCOLLOUGH, KP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (02) :471-479
[3]   CLINICAL DOSIMETRY FOR IMPLEMENTATION OF A MULTILEAF COLLIMATOR [J].
BOYER, AL ;
OCHRAN, TG ;
NYERICK, CE ;
WALDRON, TJ ;
HUNTZINGER, CJ .
MEDICAL PHYSICS, 1992, 19 (05) :1255-1261
[4]   Three dimensional conformal treatment planning with multileaf collimators [J].
Brewster, L ;
Mohan, R ;
Mageras, G ;
Burman, C ;
Leibel, S ;
Fuks, Z .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1081-1089
[5]  
British Institute of Radiology/Institution of Physics and Engineering in Medicine and Biology Joint Working Party Aird EGA British Institute of Radiology and Institution of Physics and Engineering in Medicine and Biology, 1996, BR J RADIOL S, V25
[6]   Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate [J].
Burman, C ;
Chui, CS ;
Kutcher, G ;
Leibel, S ;
Zelefsky, M ;
LoSasso, T ;
Spirou, S ;
Wu, QW ;
Yang, J ;
Stein, J ;
Mohan, R ;
Fuks, Z ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (04) :863-873
[7]   DOSE CALCULATION FOR PHOTON BEAMS WITH INTENSITY MODULATION GENERATED BY DYNAMIC JAW OR MULTILEAF COLLIMATIONS [J].
CHUI, CS ;
LOSASSO, T ;
SPIROU, S .
MEDICAL PHYSICS, 1994, 21 (08) :1237-1244
[8]  
Clarkson JR., 1941, BRIT J RADIOL, V14, P265, DOI DOI 10.1259/0007-1285-14-164-265
[9]   A MULTILEAF COLLIMATOR FIELD PRESCRIPTION PREPARATION SYSTEM FOR CONVENTIONAL RADIOTHERAPY (VOL 30, PG 707, 1994) [J].
DU, MN ;
YU, CX ;
SYMONS, M ;
YAN, D ;
TAYLOR, R ;
MATTER, RC ;
GUSTAFSON, G ;
MARTINEZ, A ;
WONG, JW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (02) :513-520
[10]   THE ACCEPTABILITY OF A MULTILEAF COLLIMATOR AS A REPLACEMENT FOR CONVENTIONAL BLOCKS [J].
FERNANDEZ, EM ;
SHENTALL, GS ;
MAYLES, WPM ;
DEARNALEY, DP .
RADIOTHERAPY AND ONCOLOGY, 1995, 36 (01) :65-74